< 9. Troubleshooting >
9-3
IM 12F5A1-01E
Err. 10: (Converter failure)
The FC400G free available chlorine analyzer has become defective. Contact Yokogawa service.
NOTE. 01: (Input Data Setting Range Exceeded)
This error appears if a set value exceeds the relevant setting range when data is set in the operation
level, setting level, or service level.
If this error message appears, the relevant data will not be updated. Set correct value again.
NOTE. 02: (Input Data Setting Error)
This error is generated if the mg/l (or ppm) value for 100% signal output falls below 1.00 mg/l (or 1.00
ppm) in the <*OUTPU> mode in the setting level.
If this error message appears, the relevant data will not be updated. Set correct value again.
NOTE. 03: (Input Data Setting Error)
This error is generated if the mg/l (or ppm) value for 100% signal output is set to a value below the
value for 50% signal output when two-line-segment output is set in service code 05 in the <*OUTPU>
mode in the setting level.
If this error message appears, the relevant data will not be updated. Set correct value again so that
the mg/l (or ppm) value for 100% signal output exceeds the value for 50% signal output.