
< 6. Operation >
6-6
IM 12F5A1-01E
6.3 Requirements for Setting Operation Data
6.3.1 Setting Items
Table 6.1 shows different Operating and Setting items. Table 6.2 shows Setting items in Service level.
Table 6.1
Setting Item List (Set Values at Shipment and Their Setting Range) -1
Setting Item
Initial Value
(at Factory Shipment)
Selection and Setting Range
Opera
-tion
Level
<DISP> (Display) Mode
• Message area display content selection
Temperature display
Sample temperature, diffusion
current, applied voltage, output
%, slope, and zero point
Setting
Level
<*OUTPU> (Output range setting) Mode
• Range 1
Setting corresponding to 50% output
1.00 [mg/L]
0.00 to 3.00 [mg/L]
Setting corresponding to 100% output
2.00 [mg/L]
0.00 to 3.00 [mg/L]
• Range 2
Setting corresponding to 50% output
1.00 [mg/L]
0.00 to 3.00 [mg/L]
Setting corresponding to 100% output
2.00 [mg/L]
0.00 to 3.00 [mg/L]
<*SET.HD> (Hold Parameter Setting) Mode
• Output hold ON / OFF selection
ON (*HLD. ON)
ON (*HLD. ON)
OFF (*HLD. OF)
• Hold output value selection
Value immediately
preceding hold (*HD.LST)
Value immediately preceding
hold (*HD. LST)
Fixed value (*HD. FIX)
• Setting of fixed value *HOLD.%
50.0 [ % ]
0.0 to 100.0 [ % ]
<*WA / CA> (Wash / Auto-zero Calibration) Mode
<*WASH> (Washing) Mode
• Washing timer ON / OFF selection
Timer OFF (* T. OFF)
Timer ON (* T. ON )
Timer OFF (* T. OFF)
• Washing interval setting (*INT H)
2.0 [ hr ]
0.1 to 24.0 [ hr ]
• Washing time setting (*WT.min)
1.0 [ min ]
0.1 to 25.0 [ min ]
• Releasing time setting (*RT.min)
5.0 [ min ]
0.1 to 25.0 [ min ]
<*AZ.CAL> (Auto-zero Calibration) Mode
• Calibration timer ON/OFF selection
Timer OFF (* T.OFF) (Note)
Timer ON / OFF
• Calibration interval setting (*INT D)
10.0 (days)
0.1 to 99.9 (days)
• Calibration setup time setting (*WT.min)
010 (min)
1 to 100 (min)
• Stabilization (settling) time setting (*RT.min)
010 (min)
1 to 100 (min)
<*PLAT> (Plateau Measurement) Mode
• Initial applied voltage setting (*INI V)
1.00 [ V ]
-1.00 to 1.00 [ V ]
• Step applied-voltage setting (*STPV)
-0.20 [ V ]
-1.00 to 1.00 [ V ]
• Step count setting (*STP.NO)
10 [ steps ]
1 to 20 [ steps ]
(Note) If you order the /AZC option, Timer ON (*T. ON) is the default.
Содержание FC400G
Страница 5: ...Blank Page ...
Страница 9: ...Blank Page ...
Страница 13: ...Blank Page ...
Страница 21: ...Blank Page ...
Страница 27: ...Blank Page ...
Страница 49: ...Blank Page ...
Страница 79: ...Blank Page ...
Страница 87: ...Blank Page ...