< 6. Operation >
6-12
IM 12F5A1-01E
(5) Code 07: Setting Applied-voltage Compensation Parameters (*DRV V Mode )
In the free available chlorine analyzer, optimum applied-voltage value changes in proportion to
the diffusion current value as shown in the plateau characteristic in Figure 6.8.
This requires applied-voltage compensation be made as expressed by the following equation.
( Applied voltage ) = ( DRV V ) + ( V/ µA ) x ( diffusion current value )
DRV V
V / µA
F6.8e.ai
Current
Optimum
applied-voltage
Applied
voltage
Figure 6.8
Plateau Characteristic of the Free Available Chlorine Analyzer
• Setting *DRV V (Start Applied-voltage Value)
Set start applied-voltage value (*DRV V) in the range of -1.50 to 1.50 V. This parameter is set to
-0.10 V at factory shipment.
Set to +0.1 V in (to decrease influence of) combined chlorine insensitive version (option).
• Setting *V/µA (Slope Value)
Set the slope value, or applied voltage change
(V/µA) per 1 µA , in the range of - 0.10 to 0.00
V/µA. This parameter is set to -0.02 V/µA at factory shipment.
Set to -0.01V/
µ
A in (to decrease
influence of) combined chlorine insensitive version (option).
(6) Code 08: Selecting Electrode Sensitivity Initialization ON / OFF (*CHANG Mode)
Set this parameter to “ 1 ” only if you wish to rewrite the slope (“ SL ”) displayed in the message
area selectable in the operation level <DISP> mode, to 1.00 (e.g., for electrode replacement).
Generally set this parameter to “ 0 ”. This allows the sensitivity relative to the sensitivity reached
when the parameter is rewritten to 1.00, to be checked using “ SL ”.
This parameter is set to “ 0 ” (initialization OFF) at factory shipment.
(7) Code 09: Washing / Auto zero calibration selection (*WA / CA Mode)
Selects either washing or auto zero calibration
For washing, select “ 0 ”, and for auto zero calibration, select “ 1 ”. Shipping-time default is “ 0 ”.
(8)
Code 10: Remote range switching ON / OFF (*R.RNGE Mode)
When input exceeds a preset voltage, the range is switched (Range can be changed).
Set “ 1 ” (ON) to enable remote range switching. Shipping-time default is “ 0 ” OFF, disabled.