< 6. Operation >
6-1
IM 12F5A1-01E
6. OPERATION
6.1 Preparation for Operation
6.1.1 Inspecting Piping and Cable Installation Condition
Confirm that the piping and wiring to the FC400G free available chlorine analyzer have been correctly
installed. For details of wiring, refer to Sec. 3.3.
6.1.2 Filling with Ceramic Beads for Electrode Polishing
Surface contamination of the indicator electrode results in measured value error. To avoid
contamination-caused error, the electrode surface is continuously polished by ceramic beads during
FC400G operation.
Ceramic beads are not filled into the beads case at factory shipment. Thus, pour the ceramic beads
provided in the accessory box into the beads case as follows before starting the FC400G.
The combined chlorine insensitive version uses glass beads.
Before operation, ceramic beads are filled into the beads case of the detector.
[ How to fill the ceramic beads ]
Unscrew two fixing screws at the bottom of the electrode mechanism block (see Figure 6.3) to remove
the measuring tank.
The ceramic beads are normally filled up to the bottommost notch of the beads case when the
indicator electrode is inserted.
F6.1e.ai
Electrode mechanism
block
Measuring tank
Fixing screws
( 2 pcs )
In maintenance position
Counter electrode
(reference electrode)
Indicator electrode
(rotating electrode)
Beads case
Ceramic beads
Beads case cover
The bottommost notch
of the beads case
Figure 6.1
Measuring Tank Parts and Beads Case