< 2. Specifications >
2-3
IM 12F5A1-01E
Ambient humidity:
5 to 95% RH (no dew condensation allowed)
Storage Temperature: -30 to 70°C
Installation:
Indoor (Outdoor use separately requires rainproof measures.) (Direct
sunlight must be avoided.)
Utility:
Power supply;
100 V AC ±10% 50/60 Hz
110 V AC ±10% 50/60 Hz
200 V AC ±10% 50/60 Hz
220 V AC ±10% 50/60 Hz
Power consumption;
Approx. 15 VA
Mounting:
2B pipe mounting or wall mounting
Weight:
Approx. 6 kg
EMC Regulatory Arrangement in Australia and New Zealand:
EN 55011 Class A, Group 1
KC Marking:
Korea Electromagnetic Conformity Standard
2.1.2 Dedicated Sampling System for (Non-Reagent Type) Free Available Chlorine
Analyzer
Function:
Used for application having high pressure and large flow rate.
Refer to GS 12A00V02-01E.
2.2
Characteristics
(% display shows value relative to the upper limit of a range.)
Repeatability:
2 %
Linearity:
±5 %
Drift:
Zero drift;
±1 % / month
Span drift;
-10 % / month
Response time:
Approx. 2 min (90% response time)
Temperature compensation error (water temperature): ±3% (Temperature compensation range : 0 to 40°C)
Ambient temperature effects:
±0.5% / 10°C
Power variation effects:
±0.5% / 10% of rated voltage
Combined chlorine effects of the combined chlorine insensitive version:
Approx. 30% of combined chlorine concentration