< 4. Parts Names and Operations >
4-4
IM 12F5A1-01E
4.2.2 Modes in Setting level
Press the
*
key to switch from Measurement to Setting mode, and press the
MODE
key to exit
Setting and revert to Measurement mode. Select the Setting mode from the following :
(1) *OUTPU :
Range setting (concentration values corresponding to 50% and 100%
output). Both range 1 and range 2 are set.
(2) *SET .HD :
Turn HOLD ON / OFF, select HOLD output (previous value, preset value).
(3) *WASH :
Select Manual or Auto Wash, select Timer ON / OFF, set wash interval,
wash time, and stabilization (settling) time. (Since the FC400G does not use
reagent, washing is not provided and wash mode is not used in the standard
instrument).
*AZ.CAL :
Set Auto-Zero calibration, select Timer ON / OFF, set wash interval,wash
time, and stabilization (settling) time.
(4) *PLAT :
Plateau characteristic measurement / data read out, initial applied voltage,
step applied voltage, set number of steps.
(5) *SERVC :
Enter service level code (refer to Sec. 4.2.3)
4.2.3 Modes in Service level
Enter the setting level using
*
key. Then, press the
NO
key to call the<*SERVC> mode and
press the
YES
key to enter the service level. The service level has the following codes required to
set various data.
Code
01 *TEMP :
Selects °C or °F
02 *FAIL :
Selects temperature compensation range exceeding fail ON/OFF.
04 *ΔT.SEC :
Sets stability determination parameter ΔT.
*ΔPPM :
Sets stability determination parameter ΔPPM.
*CT.min :
Sets stability determination parameter. CT
05 *NONLN :
Selects linear output or two-line-segment output.
07 *DRV V :
Sets applied-voltage compensation parameter Start.
*V/µA :
Sets applied-voltage compensation parameter Slope.
08 *CHANG :
Selects electrode sensitivity initialization ON/OFF.
09 *WA / CA :
Select Wash or Auto-Zero calibration.
10 *R.RNGE :
Turn ON/OFF Remote Range Selection.
Press
MODE
key to exit from service level.
(Note) If you accidentally enter codes other than 01, 02, 04, 05, 07, 08 and 10 (for example, 80),
immediately press the
MODE
key to exit from service level.