< 1. Overview >
1-2
IM 12F5A1-01E
1.1 Configuration of the Free Available Chlorine
Analyzer
The FC400G free available chlorine analyzer is broadly divided into a free available chlorine analyzer
main body (FC400G) and a sampling system: The main body consists of a detector and converter,
and the sampling system consists of sample water piping and a needle valve.
<Free Available Chlorine Analyzer Main Body>
The detector consists of a measuring tank and electrode mechanism block. The detector
electrolytically reduces free available chlorine in the sample water supplied to the measuring tank to
obtain a diffusion current in the plateau region. It also generates signals for automatic temperature
compensation of this diffusion current that varies with temperature. The measuring tank is made
of acrylic resin and contains ceramic beads in the part where a indicator electrode is inserted, for
polishing the electrode.
The electrode mechanism block incorporates an indicator electrode ( rotating electrode ), counter
electrode ( Pt 1000 Ω RTD-built-in silver chloride electrode ), and driving mechanism consisting of a
synchronous motor that rotates the indicator electrode at a constant speed, and a belt.
The converter's measuring circuit is housed together with an operation panel and external wiring
connection terminal board in a closed aluminum casing. This measuring circuit performs amplification
and computation based on two signals from the detector to generate an output signal of 1 to 5 V DC or
4 to 20 mA DC corresponding to the output range.
<Sampling System>
The FC400G sample inlet should have a pressure of 1 to 150 kPa a flow rate of 0.1 to 2.5 l/min.
Sample meeting these conditions can be directly introduced to the measuring tank. However,
measurement of sample not meeting such conditions requires installation of a sampling system to
supply the required conditions.
YOKOGAWA's dedicated sampling system (ST401G) allows measurement of sample with a pressure
of 100 to 750 kPa and flow rate of 0.1 to 10 l/min.