< 8. Maintenance >
8-8
IM 12F5A1-01E
the lengthwise direction.
(10) Place a plate that prevents friction dust from entering the electrode holder. With the two
bumps on the plate facing down, pass the plate under the driven shaft and index the two
bumps with the two holes in the plate mounted in step (11). Then, as a temporary measure,
loosely tighten the two screws that hold it in place.
(11) Mount the rotating contact on the driven shaft.
(Note) The driven shaft and rotating contact need to be in contact and make a good
electrical connection, so make sure
that they are clean. However, adherehce of lubricant for the rotating contact to the surface does not interfere with the
peformance.
(12) With the drive belt mounted, attach pulleys to the drive shaft and driven shaft.
(Note)
Install the drive shaft pulley so that the tip of the shaft can project out 0.3 to 0.7 mm from end of the shaft, so that drive
belt becomes horizontal.
(13) While adjusting the drive belt tension, further tighten the screws that were tightened
temporarily in steps (9) and (10).
Set the spring plate and the belt tension adjuster plate as Figure 8.6 shows.
Hang the hook of a spring balance (1.00 kg) on the hole in the driven shaft. Make sure that
all the temporarily tighten four screws are loose.
Draw the drive belt horizontally in the direction of the line that connects between the drive
shaft and the driven shaft, by a force of 4 N (spring balance reading of 0.4 kgf). With this
tension kept, tighten the screws completely.
Fasten the screws to set the friction dust protection plate. After the screws are fixed, remove
the belt tension adjuster.
Rotating contact
Pulleys
spring plate hook
adjuster plate
Figure 8.6
Setup of belt tension adjustment
(14) Mount indicator electrode on driven shaft, and check that indicator electrode is not
contacting the base.
(Note) If the indicator electrode is contacting the base, or if the hole in the center of the base and the center of the indicator
electrode are not aligned, loose the four screws on the motor mounting base and adjust.
(15) Remount counter electrode, reconnect wiring in place, and fix clamp.
This completes the replacement procedure. Feed power to FC400G. Check that the drive
shaft/drive belt/driven shaft rotates smoothly and there is no abnormal noise like vibration or
intermittent sound.