Error
Generation
Error Content and Causes
Measures
No.
Mode
E 1
AIR.CAL mode
Stability failure
Eliminate the causes and press the
H20. CAL mode
Even after an hour has elapsed, the
(NO)
key to perform re calibration.
measured value change does not decrease
within the set value of (mg/l)
• Sensor temperature changes.
• Dissolved-oxygen of the calibration solution changes.
• The value of A(mg/l) is not suitable.
E 2
H20. CAL mode
Zero point error failure
Clean the sensor membrane and
The zero point error exceeds the set range
recalibrate. If the error is detected again,
• Dirt sticks to the sensor membrane.
replace the electrolyte and the membrane.
• Membrane abnormality. Electrolyte degradation.
E 3
AIR.CAL mode
Slope failure
Clean the sensor membrane and
H20. CAL mode
The slope exceeds the range of 40 to 200 %
recalibrate. If the error is detected again,
MAN.CAL mode
of the theoretical value.
replace the electrolyte and the membrane.
• Dirt sticks to the sensor membrane.
• Membrane abnormality. Electrolyte degradation.
E 4
Measuring mode
Sensor membrane failure
Replace membrane
E 7
All modes
Measured temperature failure (too high)
Examine the temperatures of the
It exceeds 50.0 °C (122.0 °F).
measuring solution and sensor and the
• Measuring solution temperature is high
CODE 10 setting
• CODE 10 setting is not correct
Examine the sensor cable connection
• Sensor cable wiring failure
status.
• Temperature sensor has failed
If the temperature sensor fails (abnormal
resistance), replace the sensor.
E 8
All modes
Measured temperature failure (too low)
Examine the temperatures of the
It falls below 0 °C (or 32.0 °F).
measuring solution and sensor
• Measuring solution temperature is low
and the CODE 10 setting.
• CODE 10 setting is not correct
Examine the sensor cable connection
• Sensor cable wiring failure
status.
• Temperature sensor has failed.
If the temperature sensor fails (abnormal
resistance), replace the sensor.
E 9
All modes
Input current failure
Examine the setting for CODE 01, 02
It does not satisfy the following equation:
input current < 50 µA for galvanic sensor
< 500 nA for polarographic sensor
IM 12J6B3-E-H
8-2 Troubleshooting
8-2. Measures in the case of failure (Error) detection
If a failure is detected through the self-diagnosis of the DO402 dissolved-oxygen converter, the FAIL contact
is closed. The FAIL lamp on the operation panel lights up and an error number appears in the data display.
NOTE:
If an error is detected during configuration, the FAIL contact signal is output immediately but the error
number is displayed after that action or operation is completed.
When a FAIL contact signal is output, take measures according to Table 8-1.
Table 8-1. Measures in the case of failure (Error) detection