1-6 Introduction
1-5. System Configuration
1-5-1. Dissolved Oxygen Sensor
The DO402G can be used with a variety of sensors. Some specific programming may be necessary to
match the converter with the sensor.
a. Model DOX8SM sensor
This sensor is a galvanic sensor with a PB36 type temperature compensator and a 25 micron
membrane (1 mil).
This sensor has a nominal output current of 7.5µA in ambient air under reference conditions. In Service
code 01 selection 0 must be programmed when this sensor is used. The membrane check is inactive
with this sensor unless external solution ground is used. ( metal holder or sample line )
b. Model DO30G sensor
This sensor is essentially the same as the DOX8SM sensor, but it has a 50 micron membrane, fixed
sensor cable, Pt1000 RTD temperature compensation and integral solution ground that allows
membrane checking. The nominal output of this sensor is 3.75 µA in air under reference conditions. In
Service item”*I.CELL” selection 1 must be programmed when this sensor is used.
c. Model DO30 sensor
This sensor is specified in GS12J6K4 and it can be supplied with 50 or with 25 micron membrane.
Normally 50 micron membrane is used and no reprogramming is necessary. Otherwise Service code 02
setting must be changed from 1 to 0.
d. Ingold model 32 sensors with 12, 19 and 25 mm shaft
These sensors are available in many different configurations. The sensor is a polarographic sensor and
therefore settings of service code 01 has to be changed from 0 to 1.
The sensor output is normally between 30 and 100 nA in ambient air under reference conditions, but the
“large cathode” version has a current output between 200 and 700 nA. To utilize Slope diagnostic
features, it is best to read current output during first air calibration in “display” mode and to enter that
value in service code 01.
The temperature compensator is a NTC resistance and that can be chosen by entry of 3 in code 10.
The wiring of the sensor is different as well:
The red wire: the anode goes to terminal 18
The white wire: the cathode goes to terminal 17
The green/yellow wire: the shield goes to terminal 14 and the 2 black wires for NTC go to terminal 11
and 12.
The membrane checking feature cannot be used for these sensors, due to the construction of the
sensor.
1-5-2. The holders
a. PB30 floating ball holder
All Yokogawa sensors are compatible with the floating ball holder that is specified in GS12J6K4 and
GS12J5A1.
b. FD30 immersion fitting
The DO30 sensor is compatible with the FD30 immersion holder as specified in GS12J6K4 and with all
flow fittings and flow fitting subassemblies that are specified in GS12D7K2.
c. DOX8HS submersion type holder
The DOX8SM and the DO30G sensor are compatible with the DOX8HS holder and all PH8 model
holders.