IM 12J6B3-E-H
Parameter setting 5-19
Code
Display
Function
Function detail
X
Y
Z
Default values
Calibration Settings
20
*
∆
T.SEC
Stabilization
Stabilisation time range 10- 600 s
60 s
*
∆
mg/l
Step
DO step change range 0-50 mg/l
0.05 mg/l
or 0-300 %
21
*0.CAL
Zero cal.
Zero calibration disabled
0
0
Zero Calibration enabled
1
22
*ZERO
Zero entry
Zero range , galvanic
0 nA
Check ± 0.5 µA
No check ± 19.99 µA
Zero range, polarographic
Check ± 5 nA
No check ± 199.9 nA
*SLOPE
Slope entry
Slope: range 40- 200 %
3.75 µA (50 nA)
Minimum setting
Nominal current x 0.4
(see S.C. 01)
Maximum setting
Nominal current x 2
(see S.C. 01)