IM 12J6B3-E-H
7-2 Maintenance
7-2. Periodic maintenance for the EXA DO402 converter
The DO402 transmitter requires very little periodic maintenance. The housing is sealed to IP65 (NEMA 4X)
standards, and remains closed in normal operation. Users are required only to make sure the front window
is kept clean in order to permit a clear view of the display and allow proper operation of the pushbuttons. If
the window becomes soiled, clean it using a soft damp cloth or soft tissue. To deal with more stubborn
stains, a neutral detergent may be used.
NOTE:
Never used harsh chemicals or solvents. In the event that the window becomes heavily stained or
scratched, refer to the parts list (Chapter 9) for replacement part numbers.
When you must open the front cover and/or glands, make sure that the seals are clean and correctly fitted
when the unit is reassembled in order to maintain the housing’s weatherproof integrity against water and
water vapor. The DO measurement uses small signal currents and may otherwise be prone to problems
caused by exposure of the circuitry to condensation.