10. APPENDIX
Adjustable limits and defaults for settings in Commissioning Level
10-1. Setpoint
10-2. Range
NOTE:
The range has a minimum span of: 1.0 mg/l, 1.0 ppm or 10 %sat.
In case of a temperature output 25 ºC or 50 ºF
10-3. Hold
10-4. Wash
Appendix 10-1
IM 12J6B3-E-H
Variable
Default
Lower limit
Upper limit
*OX1
19.5 mg/l
0.0 mg/l
50.0 mg/l
0.0 ppm
50.0 ppm
0.0 %sat
300 %sat
*OX2
1.0 mg/l
0.0 mg/l
50.0 mg/l
0.0 ppm
50.0 ppm
0.0 %sat
300 %sat
*OX3
10.0 mg/l
0.0 mg/l
50.0 mg/l
0.0 ppm
50.0 ppm
0.0 %sat
300 %sat
*OX4
10.0 mg/l
0.0 mg/l
50.0 mg/l
0.0 ppm
50.0 ppm
0.0 %sat
300 %sat
*Sp.mA2
10.0 mg/l
0.0 mg/l
50.0 mg/l
0.0 ppm
50.0 ppm
0.0 %sat
300 %sat
* T1
25.0 ºC/77 ºF
0.0 ºC/30 ºF
50 ºC/130 ºF
* T2
25.0 ºC/77 ºF
0.0 ºC/30 ºF
50 ºC/130 ºF
* T3
25.0 ºC/77 ºF
0.0 ºC/30 ºF
50 ºC/130 ºF
* T4
25.0 ºC/77 ºF
0.0 ºC/30 ºF
50 ºC/130 ºF
Variable
Default
Lower limit
Upper limit
Range 1:
0% = 0.00
0.0 mg/l
50.0 mg/l
Process
100% = 19.99
0.0 ppm
50.0 ppm
0.0 %sat
300 %sat
Range 2:
0% = 0.00
0.0 mg/l
50.0 mg/l
Process
100% = 19.99
0.0 ppm
50.0 ppm
0.0 %sat
300 %sat
Temperature
0% = 0 ºC / 0 ºF
0.0 ºC / 30 ºF
50 ºC / 130 ºF
100% = 50 ºC / 122 ºF
Variable
Default
Lower limit
Upper limit
mA fix 1
10.00 mA
0 or 4 mA
20 mA
mA fix 2
10.00 mA
0 or 4 mA
20 mA
Variable
Default
Lower limit
Upper limit
Interval
6 hours
0.1 hours
36.0 hours
Wash time
0.5 minutes
0.1 minutes
10.0 minutes
Recovery time
0.5 minutes
0.1 minutes
10.0 minutes