Specifications 2-3
2-2. Operating specifications
A. Performance : DO (at t process = 25 °C)
- Linearity
:
≤
0.03 mg/l ± 0.02 mA
- Repeatability
:
≤
0.03 mg/l ± 0.02 mA
- Accuracy
:
≤
0.05 mg/l ± 0.02 mA
B. Performance : Temperature (Pt1000,
2k2NTC, 22kNTC)
- Linearity
:
≤
0.3 ºC ± 0.02 mA
- Repeatability
:
≤
0.1 ºC ± 0.02 mA
- Accuracy
:
≤
0.3 ºC ± 0.02 mA
Performance : Temperature (Pt100)
- Linearity
:
≤
0.4 ºC ± 0.02 mA
- Repeatability
:
≤
0.1 ºC ± 0.02 mA
- Accuracy
:
≤
0.4 ºC ± 0.02 mA
C. Response time
0- 90%
: 10 s
NOTE:
The specifications are expressed with simulated
inputs, because the DO402G can be used with
many different sensors with their unique
characteristics.
D. Ambient operating temperature
: -10 to +55
o
C (10 to 131 ºF)
Excursions to -30 °C do not
influence the current output
function, and excursions to
+70 °C are acceptable too.
E. Storage temperature
: -30 to +70
o
C (-20 to 160 ºF)
F. Humidity
: 10 to 90% RH non-
condensing
G. Housing
Case
: Cast aluminium with
chemically resistant coating
Cover
: flexible polycarbonate window.
Case color
: off-white
Cover color
: moss green.
Cable entry
: via six
1
/
2
” polyamide glands.
Cable terminals: for up to 2.5 mm
2
finished
wires.
Protection
: weather resistant to IP65 and
NEMA 4X standards.
Mounting
: Pipe, wall or panel, using
optional hardware.
H. Data protection
: Non volatile memory for
configuration and logbook,
and lithium battery for clock
support.
I. Watchdog timer
: Checks microprocessor
J. Automatic safeguard
: Return to measuring mode
when no keystroke is made
for 10 min.
K. Power interruption
: Less than 50 milliseconds no
effect. More than 50
milliseconds reset to
measurement.
L. Operation protection
: 3-digit programmable
password.
M. Regulatory compliance
- EMC
: meets council directive
89/336/EEC
- Emission
: meets EN 55022 Class A
- Immunity
: meets EN 50082-2
- Low voltage
: meets council directive
73/23/EEC
- Installation
: Designed for installation
conforming to IEC 1010-1.
Category II.
2-3. Model and suffix codes
IM 12J6B3-E-H
Model
Suffix
Option Description
code
code
DO402G
.................................... Converter for dissolved
oxygen
-E ................................ Always E
Supply
-1 ........................... 115 Volts 50/60 Hz
voltage
-2 ........................... 230 Volts 50/60 Hz
-4 ........................... 24 Volts DC
-5 ........................... 100 Volts 50/60 Hz
Instruction manual -E .................... English language*
Options
/U......... Pipe and wall mounting
hardware
/PM...... Panel mounting hardware
/Q ........ Quality certificate
/SCT .... Stainless steel tag
* For other languages contact local sales office
Содержание EXA DO402G
Страница 1: ...Instruction Manual Model DO402G Converter for Dissolved Oxygen IM 12J6B3 E H 5th Edition YOKOGAWA...
Страница 2: ......
Страница 6: ...IM 12J6B3 E H...
Страница 16: ...IM 12J6B3 E H...
Страница 60: ...IM 12J6B3 E H...
Страница 74: ...IM 12J6B3 E H...
Страница 76: ...IM 12J6B3 E H...