Parameter setting 5-15
IM 12J6B3-E-H
Code
Display
Function
Function detail
X
Y
Z
Default values
Parameter Specific functions
01
*S. TYPE
Sensor type
Galvanic
0
0
Polarographic
1
*I CELL
Sensor output
3.75 µA for 50 micron type
0
0
Galvanic
7.50 µA for 25 micron type
1
User defined
9
3.75 µA
*I CELL
Sensor output
50 nA
0
0
Polarographic
User defined
9
50 nA
02
*CHECK
Sensor check
Zero check: disabled
0
0.1.0
Zero check: enabled
1
± 0.5 µA
Slope check: disabled
0
min. 40%
Slope check: enabled
1
max. 200%
Membrane check: disabled/only for Galvanic
0
Membrane check: enabled/only for Galvanic
1
03
NOT USED
04
*SAL.TY
Salinity comp.
Compensation disabled
0
0
Compensation enabled
1
*[Cl-]
NaCl conc.
Concentration in g/kg water:
0.0
range 1 to 99.9 g/kg
05-09
Not used
Содержание EXA DO402G