![YOKOGAWA EXA DO402G Скачать руководство пользователя страница 14](http://html1.mh-extra.com/html/yokogawa/exa-do402g/exa-do402g_instruction-manual_3467515014.webp)
2-2 Specifications
IM 12J6B3-E-H
J. Contact outputs
- General
: Four (4) SPDT relay contacts
with LED indicators. For S1,
S2, and S3, the LED is on
when relay is powered.
NOTE:
For S4 (FAIL) LED lights when
power is removed (Fail safe).
Contact outputs configurable
for hysteresis and delay time.
- Switch capacity
: Maximum values 100 VA,
250 V AC, 5 Amps.
Maximum values 50 Watts,
250 V DC, 5 Amps.
- Status
: High/Low process alarms,
selected from process
parameters and temperature.
Contact output is also
available to signal “Hold
Active”
- Control function
: On/Off
PI pulsed
Proportional duty cycle control
with integral term.
PI frequency
Proportional frequency control
with integral term. In addition
wash cleaning control signal
on S3, and FAIL alarm for
system and diagnostic errors
on S4
K. Contact input
: Remote wash cycle start.
L. Power supply
: 230 V AC ±15%, 50/60 Hz.
Max. consumption 10 VA.
115 V AC ±15%, 50/60 Hz.
Max. consumption 10 VA.
100 V AC ± 15% 50/60 Hz.
Max. consumption 10 VA.
24 V DC -20% / +30%
Max. consumption 10 Watts.
M. Shipping details
: Package size w x h x d
290 x 225 x 170 mm.
11.5 x 8.9 x 6.7 in.
Packed weight approx.
2.5 kg (5lb).
Содержание EXA DO402G