46
IM CA310-02EN
CA330
RTD Source/Measurement Specifications
RTD
Coefficient
Accuracy (1 year)
Excitation
Current Input
Range
Standard, Citation
Temperature Range
Measurement
Accuracy [°C]
Source
Accuracy [°C]
PT100
3851
-200.0°C ≤ t < 0.0°C
0.3
0.3
0.1 to 3 mA
IEC60751
JIS C 1604
0.0°C ≤ t ≤ +800.0°C
0.3+t x 0.033% 0.3+t x 0.033%
3850
-200.0°C ≤ t < 0.0°C
0.3
0.3
0.1 to 3 mA
JIS C 1604 1989
(Pt100)
0.0°C ≤ t ≤ +630.0°C
0.3+t x 0.033% 0.3+t x 0.033%
3916
-200.0°C ≤ t < 0.0°C
0.3
0.3
0.1 to 3 mA
JIS C 1604 1989
(JPt100)
0.0°C ≤ t ≤ +510.0°C
0.3+t x 0.033% 0.3+t x 0.033%
3926
-200.0°C ≤ t < 0.0°C
0.3
0.3
0.1 to 3 mA
Minco Application Aid #18
0.0°C ≤ t ≤ +630.0°C
0.3+t x 0.033% 0.3+t x 0.033%
PT200
3851
-200.0°C ≤ t < 0.0°C
0.3
0.3
0.05 to 0.8 mA IEC60751
0.0°C ≤ t ≤ +630.0°C
0.3+t x 0.050% 0.3+t x 0.050%
PT500
3851
-200.0°C ≤ t < 0.0°C
0.4
0.4
0.05 to 0.6 mA IEC60751
0.0°C ≤ t ≤ +630.0°C
0.4+t x 0.033% 0.4+t x 0.033%
PT1000
3851
-200.0°C ≤ t < 0.0°C
0.2
0.2
0.05 to 0.6 mA IEC60751
0.0°C ≤ t ≤ +630.0°C
0.2+t x 0.033% 0.2+t x 0.033%
Cu10
427
-100.0°C ≤ t < +260.0°C
1.5
1.5
0.1 to 3 mA
Minco Application Aid #18
Ni120
627
-80.0°C ≤ t < +260.0°C
0.2
0.2
0.1 to 3 mA
Minco Application Aid #18
Specifications