Installation Conditions
Install the instrument in a place that meets the following conditions.
• Operating Altitude and Ambient Temperature and Humidity
• Use the instrument in the following environment.
• Ambient temperature: –10°C to 55°C
• Ambient humidity: 20% RH to 80% RH; no condensation
• Operating altitude: 2000 m or less
Note
• To ensure accurate use, operate the instrument within 23°C ± 5°C.
• When using the instrument in a place where the ambient humidity is 30% or less, take measures to prevent static electricity such as
using an anti-static mat.
• Condensation may occur if the instrument is moved to another place where the ambient temperature or humidity is higher, or if the
temperature changes rapidly.
In such cases, before you use the instrument, allow it to adjust to the surrounding temperature for at least an hour.
Do not install the instrument in the following places.
• In direct sunlight or near heat sources
• In an environment with excessive amounts of soot, steam, dust, or corrosive gas
• Near strong magnetic field sources
• Near noise sources, such as high-voltage equipment or power lines
• In an environment subject to large levels of mechanical vibration
• On an unstable surface
• In an environment where ignition or explosion may occur, such as in explosive gas