40
IM CA310-02EN
Using Terminal B (Banana Terminal), Reference Junction Compensation Based on External RJ Sensor (sold separately)
(t: measurement/source temperature)
Thermocouple
Accuracy (1 year)
Specification
Temperature Range
Source Accuracy [°C]
Measurement
Accuracy [°C]
K
-200.0°C ≤ t < 0.0°C
0.7+|t| x 0.4%
0.7+|t| x 0.4%
EC60584-1
JIS C1602
0.0°C ≤ t < +500.0°C
0.7
0.7
+500.0°C ≤ t ≤ +1372.0°C
0.7+(t-500) x 0.03%
0.7+(t-500) x 0.03%
E
-250.0°C ≤ t < -200.0°C
1.3+(|t|-200) x 5.0%
1.3+(|t|-200) x 5.0%
IEC60584-1
-200.0°C ≤ t < 0.0°C
0.7+|t| x 0.3%
0.7+|t| x 0.3%
0.0°C ≤ t < +500.0°C
0.7
0.7
+500.0°C ≤ t ≤ +1000.0°C
0.7+(t-500) x 0.02%
0.7+(t-500) x 0.02%
J
-210.0°C ≤ t < 0.0°C
0.7+|t| x 0.3%
0.7+|t| x 0.3%
IEC60584-1
0.0°C ≤ t ≤ +1200.0°C
0.7+t x 0.02%
0.7+t x 0.02%
T
-250.0°C ≤ t < -200.0°C
1.7+(|t|-200) x 5.0%
1.7+(|t|-200) x 5.0%
IEC60584-1
-200.0°C ≤ t < 0.0°C
0.7+|t| x 0.5%
0.7+|t| x 0.5%
0.0°C ≤ t ≤ +400.0°C
0.7
0.7
N
-200.0°C ≤ t < 0.0°C
0.8+|t| x 0.5%
0.8+|t| x 0.5%
IEC60584-1
0.0°C ≤ t ≤ +1300.0°C
0.8
0.8
L
-200.0°C ≤ t < 0.0°C
0.7+|t| x 0.15%
0.7+|t| x 0.15%
DIN 43710
0.0°C ≤ t ≤ +900.0°C
0.7
0.7
U
-200.0°C ≤ t < 0.0°C
0.7+|t| x 0.3%
0.7+|t| x 0.3%
DIN 43710
0.0°C ≤ t ≤ +600.0°C
0.7
0.7
R
-20.0°C ≤ t < 0.0°C
1.4+|t-100| x 0.5%
1.4+|t-100| x 0.5%
IEC60584-1
0.0°C ≤ t < +100.0°C
1.4+|t-100| x 0.5%
1.4+|t-100| x 0.5%
+100.0°C ≤ t ≤ +1767.0°C
1.4
1.4
S
-20.0°C ≤ t < 0.0°C
1.4+|t-100| x 0.5%
1.4+|t-100| x 0.5%
IEC60584-1
0.0°C ≤ t < +100.0°C
1.4+|t-100| x 0.5%
1.4+|t-100| x 0.5%
+100.0°C ≤ t ≤ +1768.0°C
1.4
1.4
Specifications