34
IM CA310-02EN
Specifications
CA310
Measurement Specifications
Function
Range
Measurement Range
Resolution
Accuracy (1 year)
Notes
of Reading
Offset
Voltage
500 mV
0.00 to ±550.00 mV
10 μV
0.015%
50 μV
Input resistance: Approx.1 MΩ
5 V
0.0000 to ±5.5000 V
0.1 mV
0.015%
0.5 mV Input resistance: Approx.1 MΩ
30 V
0.000 to ±33.000 V
1 mV
0.015%
5 mV
Input resistance: Approx.1 MΩ
50 V
0.000 to ±55.000 V
1 mV
0.015%
5 mV
Input resistance: Approx.1 MΩ
Current
20 mA
0.000 to ±24.000 mA
1 μA
0.015%
3 μA
Input resistance: 10 Ω or lower
50 mA
0.000 to ±60.000 mA
1 μA
0.015%
3 μA
Input resistance: 10 Ω or lower
LOOP
POWER
20 mA
0.000 to +24.000 mA
1 μA
0.015%
3 μA
Supply voltage:
24 V±1 V (when HART communication resistance is off. Maximum load current: 24 mA)
24 V±6 V (when HART communication resistance is on. Maximum load current: 20 mA)
At the ambient temperature (Ta) of +23°C±5°C.
For the accuracies of other ambient temperature ranges (Ta<18°C, Ta>28°C), add the temperature coefficient 0.005% of Range/°C.
CMRR
Approx. 120 dB (50/60 Hz)
NMRR
Approx. 60 dB (50/60 Hz)
Measurement terminal maximum input
Voltage terminals 50 VDC
Current terminals 50 mADC
Current terminal input protection
PTC protection
Measurement display update rate
Approx. 1 s
Measurement terminal voltage to ground 50 Vpeak