45
Voltage Source/Measurement Specifications
Range
Measurement and Source Range
Resolution
Accuracy (1 year)
Notes
of Setting
of Reading
Offset
90 mV
–11.000 to +99.999 mV
1 μV
0.015%
10 μV
Maximum output current: 1 mA
At the ambient temperature (Ta) of 23±5°C.
For the accuracies of other ambient temperature ranges (Ta<18°C, Ta>28°C), add the temperature coefficient 0.005% of Range/°C.
Common Measurement Section Specifications
Maximum input across terminals: 42 Vpeak
Maximum input to ground: 42 Vpeak
Common Source Section Specifications
Output resistance: 40 mΩ or less
Output response: 300 ms or less
Maximum load: C≤0.1 μF, L≤10 mH
Maximum input across terminals: 42 Vpeak
Maximum input to ground: 42 Vpeak
Thermocouple Accuracy Equation
Accuracy with respect to the measurement or source temperature (t) is expressed as constant or a linear equation of t.
Example: Accuracy for Thermocouple K (terminal A) measurement value of 1000.0°C = ±(0.5 + (1000.0-500)×0.02%)°C = ±0.6°C
Specifications