IM CA310-02EN
33
Problems and Solutions
Reference
Page
“Err60” appears at power-on. This indicates that the settings have been initialized due to an error in the
setting information stored in the CA310, CA320, or CA330.
If the error appears every time at power-on, repair is necessary.
-
“Err61” and “Err63” appear at
power-on.
There is an error in the calibration data stored in the CA310, CA320, or
CA330.
If the error appears every time at power-on, repair is necessary.
-
“Err30” appears when the
span 0% or span 100% value
is set.
If the span 0% value is set greater than or equal to the span 100% value,
an error occurs. Set to a value less than the span 100% value.
If the span 100% value is set less than or equal to the span 0% value, an
error occurs. Set to a value greater than the span 0% value.
15
Recommended Replacement Parts, Consumable Parts, and Calibration
Recommended Replacement Parts, Consumable Parts, and Calibration
Recommended Replacement Parts and Consumable Parts
YOKOGAWA guarantees the instrument for the period and under the conditions of the product warranty.
All parts of this instrument are covered by the one-year warranty.
Calibration
To ensure accurate measurement, we recommend periodic calibration.
Recommended calibration period: 1 year
Disposing of the Instrument
When disposing of the instrument, follow the laws and ordinances of your country or region.
Troubleshooting, Maintenance, and Inspection