<6. PARAMETER DESCRIPTION>
6-14
IM 01E30D01-01EN
(11) Item M (Menu M): Automatic Zero-Adjustment Function Setting items
Menu M contains items that are relevant to automatic zero adjustment.
Item
Name
R/W
Data range
Units
Position
of decimal
point
Default value
(*): Indicated item
Description
Display unit
(BRAIN)
Display unit
/BRAIN
M00 Adjustment
(ADJUSTMENT)
M10 Auto Zero Exe
(AUTOZERO EXE)
W
No Execution
Execution
No Execution
Selects whether or not
automatic zero adjustment is
carried out. Linked with B50.
M11 Autozero Time
(AUTOZ TIME)
W
1 to 900
s
0
450 s
Sets execution time for
automatic zero adjustment
M12 Flow Zero(IEL)
(FL ZERO(IEL))
W
-99.999 to 99.999
3
0.000
Displays the result of automatic
zero adjustment (excitation
current (Low))
M13 Flow Zero(IEM)
(FL ZERO(IEM))
W
-99.999 to 99.999
3
0.000
Displays the result of automatic
zero adjustment (excitation
current (Middle))
M14 Flow Zero(IEH)
(FL ZERO(IEH))
W
-99.999 to 99.999
3
0.000
Displays the result of automatic
zero adjustment (excitation
current (High))
M15 Flow Span Adjust
(FL SPAN ADJ)
W
0.4000 to 2.0000
4
1.0000
Sets adjustment factor of flow
span
M20 Adjustment 4mA
(ADJUST 4mA)
W
-10.00 to 10.00
%
2
0.00 %
Sets adjustment for 4mA output
M21 Adjustment 20mA
(ADJUST 20mA)
W
-10.00 to 10.00
%
2
0.00 %
Sets adjustment for 20mA
output
M60 —
(SELF CHECK)
R
Good
Error
Read Section 6.5 “Alarm
Functions”.
(12) Item N (Menu N): Loop Test Setting items
Menu N contains items that are relevant to the execution of loop testing.
Item
Name
R/W
Data range
Units
Position
of decimal
point
Default value
(*): Indicated item
Description
Display unit
(BRAIN)
Display unit
/BRAIN
N00 Test
(TEST)
N10 Test Mode
(TEST MODE)
W
Normal
Test
Normal
Selects whether mode will be
set to “Normal” or “Test”.
N11 Test Output Value
(TEST OUT VAL)
W
-1 to 103
%
0
0%
Sets the test output value.
N12 Test DO
(TEST DO)
W
Open(Off)
Closed(On)
Pulse
Open(Off)
Selects the test condition for
DO terminal.
N15 Test Time
(TEST TIME)
W
10min
30min
60min
3h
6h
12h
30min
Select the holding time of the
test mode
N20 Test Iex
(TEST IEX)
W
Normal
Test (Iex OFF)
Normal
Selects the test of excitation
current
N30 Flow Tube
(FLOW TUBE)
W
Detector
Calibrator
Detector
Selects type of operating
flowtube
N40 Average Execution
(AVG EXEC)
W
Stop
Start
Stop
Selects whether the execution
for measurement of averaged
flowrate is occurrence
N41 Average Timer
(AVG TIMER)
W
1 to 600
s
0
30 s
Sets totalization time for
measurement of averaged
flowrate
N42 Average Flow
(AVG FLOW)
R
-999999 to 999999 C40/C41
0 to 5
0.00000 m/s
Displays the averaged flowrate
N60 —
(SELF CHECK)
R
Good
Error
Read Section 6.5 “Alarm
Functions”.