7.1 Specifications
IM 735201-01E
7-3
Combined Calibration Specifications
The sensor head (735221) and TB200 Optical Powermeter (735201) are calibrated
together. If you selected the combined calibration option at the time of purchase,
the electrical/optical characteristics of the sensor head are as given below (suffix
code -CA1 or -CA3). The sensor head calibration data is written to the TB200
Optical Powermeter (735201), and the model of the sensor head becomes that
of the main unit (735201).
Electrical/Optical Characteristics (With the Combined Calibration Option)
Item
Specification
Wavelength range
400 to 850 nm
Light-receiving element
Si-PD
Received light power
range
1
µ
W (-30 dBm) to 100 mW (+20 dBm) Note 1)
Max. light receiving level
+20 dBm (100 mW) Note 1)
Max. power density
5 mW/mm
2
Note 1)
Uncertainty
at
reference
conditions
±2.5
%(
405 nm
)
±3.0 %
(
660 nm 785 nm
)
Note 2)
Input type
Spatial light
Accessories
TB200 Utility CD Note 3)
Sensor protective cap
Note 1)
Condition:
λ
= 405 nm
Note 2)
Reference conditions:
(1) Reference temperature: 23 °C ±3 °C
(2) Reference wavelength:
λ
= 405 nm (Add 0.5% when the wavelength
is in the range of 400 to 420 nm.)
660 nm calibration option:
λ
= 660 nm
785 nm calibration option:
λ
= 405 nm
(3) Reference power: 1 mW
(4)
Reference beam shape: Distribution: Gaussian distribution,
Radiated NA: 0.2, diffused light (50GI fiber output)
(5) Spectral width: 1 nm or less
(6) Light receiving position: Mechanical center
(7) Wavelength setting error: Within _0.5 nm
(8) Not including secular changes of measuring equipment
(9) Uncertainty inclusion coefficient: k = 2
* Uncertainty when only sensor head is sold. For details on
uncertainty when the integrated calibration option is applied, refer
to the “Remarks” column of the Model and Suffix Code table.
Note 3) Compensation values for this sensor head are provided in the TB200
Utility CD.
These performance values are for when this data is uploaded to the
TB200 Optical Power Meter for use.