3.3 Absolute Value Measurement
IM 735201-01E
3-11
5. You can set an arbitrary CAL value in 0.01 dB steps.
After setting the display to the lower right of “
CAL
” to “
USr
”, press the
ENTER
key. The CAL detail setting screen appears.
6. Press
the
UP
or
DOWN
key. The CAL value changes in steps of 0.01 dB.
The available setting range is -10 dBm to +10 dBm
Figure 3-16
Example of the CAL value detail setting screen
7. Press
the
ENTER
key. The instrument returns to normal measurement
mode. The CAL detail setting is finalized.
Figure 3-17
Example of the normal measurement mode screen when
the CAL setting is enabled