![YOKOGAWA 735201 Скачать руководство пользователя страница 21](http://html.mh-extra.com/html/yokogawa/735201/735201_user-manual_911162021.webp)
1.2 List of Functions
IM 735201-01E
1-7
1. Moving
average
Averages measured values from a determined number of measurements.
The average value is updated at all times as the most recent values are
loaded and the oldest ones are discarded.
2. The settings that are saved by the Resume function are as follows.
●
Absolute value/relative value setting mode
●
Range AUTO/HOLD
●
Relative measurement value
●
Range info. (only with range hold)
●
Measurement wavelength
●
NA correction value ON/OFF
●
Last set detail wavelength
●
Last set NA correction value
●
Displayed units
●
CAL setting ON/OFF
●
Averaging ON/OFF
●
Last set CAL value
●
Turning MAX Hold ON/OFF
Power save
(only when using a
battery)
When running on dry cells, the power turns OFF automatically if
a key is not pressed within ten minutes.
Battery alarm
(only when using a
battery)
The battery indicator blinks when the battery runs low.
Resume
2
The settings that were active when the power was turned OFF
are saved and restored the next time the power is turned ON (if
shutdown successfully).
The function is disabled when the battery alarm indicator is
displayed. In this case, the instrument saves the settings that
were last active when the power was turned OFF normally.
Analog output
3
Outputs an analog voltage for each range according to the
measured value.
USB communication
Lets you edit settings and acquire measured values via USB
(keys cannot be used for control when using this function).
Meas val 1
Meas val 2 Meas val 3
・・・
Meas val N Meas val N+1 Meas val N+2
Next, avg. over this area
Then this area, and so on
Avg. over this area
N : No. of measurements
(