5.1 Uploading Sensor Specific Data
IM 735201-01E
5-2
4. Click
File Open
.
The Calibration File Select screen opens.
5.
Select a data file from the Calibration Data folder. Click
Open
.
6. Click
Data Upload
. The upload finishes.
7. Click
Exit
.
Remove the USB cable. The instrument enters normal measurement mode
(use a safe procedure for removing the USB connection from the PC side).
Note
・
・
The uploaded correction data is unique to the sensor head.
When replacing it with a different sensor head, you must upload the
correction data for the new sensor head.