6.1 Error Messages
IM 735201-01E
6-2
Resume Error
If the Resume value was not saved successfully, the
Resume error appears the next time the instrument is
started.
To cancel a resume error, press the
ENTER
key or turn the
power OFF and restart.
The normal measurement
mode screen appears.
CAL Value Setting Error
This is displayed if calibration data was not input or was lost.
Upload sensor specific data (calibration data) via USB (see
section 5.1).
Note that if you selected the combined calibration option at
the time of purchase (suffix code -CA0), the sensor specific
data (calibration data) has already been loaded into the
TB200 Optical Powermeter.
Figure 6-6
CAL value
setting error
Figure 6-5 Resume error