![YOKOGAWA 735201 Скачать руководство пользователя страница 60](http://html.mh-extra.com/html/yokogawa/735201/735201_user-manual_911162060.webp)
Chapter 7 Specifications
IM 735201-01E
7-1
7.1 Specifications
TB200 Optical Power Meter: 735201
Environmental Conditions
Item
Environmental Condition
Operation-guaranteed
temperature/humidity
conditions
Body: 5 to 40°C (ambient temperature), 20 to 80% (no
condensation)
Storage
temperature/humidity
-20 to +60°C (ambient temperature), 20 to 80% (no
condensation)
Electrical Specifications
Item
Specification
Display
7-segment, 4-digit, w/ backlight
Display resolution
0.01 dB (When W unit is selected, floating point w/
4 digits past decimal point)
Unit display
Absolute value: dBm, mW, µW, nW
Incremental value: dB
Wavelength setting range
400 to 850 nm
Wavelength sensitivity
compensation increment
1 nm
Range selection
AUTO
/
HOLD
NA compensation range
0.500 to 2.000 (0.001 increments)
Optical power display range
1 µW (-30 dBm) to 100 mW (+20 dBm)
Measurement interval
Approx. 100 msec
Backlight
Lights when backlight key is pressed, and goes out when
key is pressed again.
Analog output
0 to 2 V connector: UM connector (made by Hirose
Electric)
Interface
USB (B connector)
Sensor head
Model: 735221 (Model name of sensor head becomes
735201 when -CA1 or -CA3 integrated calibration is
selected. However, its performance is the same.)
Power supply
AC adapter (rated input voltage: 100 to 240 V, 50/60 Hz,
0.24 to 0.13 A) 7 VA AA alkali dry cell (operation time:
approx. 24 hours)
Accessories
User’s Manual, AC adapter, UM connector protective cap
Safety, EMC
Note)
Safety: EN61010-1,
(Out of conformance since December 1, 2010.)
EMC: EN61326-1 ClassA, Table2 (for use in industrial
locations)Pollution degree 2, Overvoltage category II
Note) Applies to products with CE marks. For all other products, contact your nearest
YOKOGAWA dealer.