IM 735201-01E
4-4
4.3 List of Commands
The following describes the commands available on the TB200.
The TB200 offers ASCII character strings in addition to input and output commands.
Also, the return values for setting commands and query commands are separate, so
there is no need to send commands and queries separately.
Command Notation Conventions and Meanings
1) Commands begin with a colon ( : ).
2) <wsp> means,
enter a space.
3) <Value> means,
enter value here
.
4) ON|OFF means,
enter
e
ither ON or OFF
.
Command List
No.
Function
Command
Description
Parameters
Response
1 Data
request
:READ:POW?
Returns the
current
measured value.
None
Returns the current
measured value in Float
format. The units are the
same as those during
setting.
ERR#: See error list.
2 Data
request with
status
:READ:POW:
STAT?
Returns the
current
measured value
and range
status.
None
Returns the current
measured value in Float
format.
<Range status>
Over Range
Under Range
In Range
Average
ERR#: See error list.
3 Max
value
data
request
:READ:MAX?
Returns the
current
maximum
measured value.
None
Returns the current
maximum measured value
in Float format. The units
are the same as those
during setting.
ERR#: See error list.
4 Wavelength
setting
:SENS:POW:
WAV<wsp>
<Value>
Sets the
wavelength
400-850
(1 step)
Units: nm
OK: Setting complete
ERR#: See error list.
5 Check
the
set
wavelength
:SENS:POW:
WAV?
Returns the set
wavelength.
None
Set as a three digit integer.
The units are fixed at nm.
ERR#: See error list.
6 NA
correction
function
enable/
disable
setting
:SENS:CORR
:NA:STAT<ws
p>
ON|OFF
Turns NA
correction
function ON and
OFF
ON:
Enabled
OFF:
Disabled
OK: Setting complete
ERR#: See error list.