3.3 Absolute Value Measurement
IM 735201-01E
3-8
3.3.3 Fixing the Measuring Range (Range Hold)
You can fix the measuring range (range hold). This is useful when using the
analog out function.
1. Perform the operation from the initial screen in normal measurement mode
(Figure 3-1).
2. Press
the
RANGE
key. The “
HOLD
” display and current range value blink
for five seconds.
Figure 3-11
Example of the range value display during range hold
3. After blinking, “
HOLD
” is displayed, and the range value is no longer
displayed. The fixed range setting is finalized. The instrument enters range
hold mode.
Figure 3-12
Example of the display when in range hold mode
To Change the Range Value
4. Press
the
UP
or
DOWN
key while in range hold mode. The range goes up
or down.
5. The range blinks for five seconds in the area shown in Figure 3-11, then the
original screen returns (if REF is set, the REF value is displayed again).
Note
・
・
If the measuring range is fixed and the incoming light exceeds the
measuring range, the measured value blinks. If the upper limit of the
range is exceeded, “
▲
” is displayed. If the lower limit is not exceeded
“
▼
” is displayed.
Press the
RANGE
key again to return to AUTO mode.
Blinks for 5 sec. in synch. with the range value display,
then stops blinking.
Range value display
Displays the current range value for
5 sec. (blinks)