34
8. TROUBLESHOOTING
Troubleshooting Guide
Symptom
Check
Unit does power on.
Check to make sure power cable is securely inserted into the
receptacle on back of unit
Check to see whether power failure is in progress.
Check whether water/oil level is low.
Circuit breaker (main power switch)
trips.
Check whether unit exterior is wet.
Short-circuit (call for service.)
Er06 appears in display.
Check whether temperature setting is lower than bath
temperature.
Temperature does not come down.
Check whether temperature setting is lower than bath
temperature.
Temperature reading too high/low.
Temperature may not have stabilized yet.
Check to see whether an incorrect calibration offset value
has been entered. (See P.27)
Check to see whether bath temperature sensor has
contaminant buildup, and whether it is properly contacting
bottom surface of bath container.
If problems persist, turn off main power switch immediately, disconnect power cable and call for
service.
Содержание BM500
Страница 2: ......