18
4. OPERATION PROCEDURES
Constant Temperature Mode
Constant Temperature mode is a general use, continuous operation mode (started/stopped manually).
Constant temperature
operation procedures
1. Turn ON main power switch.
When power switch is turned ON, display shows initial values
for about four seconds before entering standby, which shows
current bath temperature in display.
See
“Display Characters” (P.14).
2. Set temperature
Set the desired temperature using
△▽
.
Press
△
or
▽
. Temperature setting is shown. Press
△
to
increase value. Press
▽
to decrease value.
After three seconds, display will revert to current bath
temperature (standby).
CAUTION: If hot water is added during operation, bringing
temperature to 40°C or more beyond temperature setting, the unit
interprets this as an overheating event. Power is cut to the heater,
and error 6 is displayed. If bath water level becomes too low while
unit is running, cancel operation before adding.
3. Start operation
Press and hold START/STOP.
Operation begins. The operation and HEATER lamps light and
temperature begins to build.
4. Stop operation
Press and hold START/STOP.
Operation stops. The operation and HEATER lamps go out.
The display reverts to standby (current bath temperature).
Correcting or viewing settings.
Use
△▽
at any time to correct or view settings.
Press either
△
or
▽
once to view temperature setting. Display will automatically revert to temperature
reading after 4 seconds. To change the setting, continue operating
△▽
until desired temperature is
entered. Unit will continue operation under new temperature setting.
Temperature reading/setting display:
Current bath temperature is displayed in standby.
Содержание BM500