26
4. OPERATION PROCEDURES
Keypad Lock Function
Using keypad lock function
This function locks the keypad, preventing unwanted interruptions or
setting changes during unit operation. Use the TIMER key to set or
cancel this function.
①
Press and hold TIMER for about four seconds. Select [LocK] to
enter menu.
②
[LocK] and [off] are displayed alternately. Select [on] using
△
to
lock keypad.
③
Select [off] using
▽
to unlock.
When function is set to [on], all keys except TIMER are locked.
Bath Setting Function
This function is available only when RE600/800 rotary evaporator unit is connected.
Connect the RE unit vacuum controller and BM/BO bath unit interfaces using the connection cable
supplied with the RE600/800 unit. Refer to the VR600/VR800 instruction manual for details.
Bath operation setting
Using signals from/to the RE unit vacuum controller, this function
stops bath completely or retains heat in bath when the rotary
evaporator completes an operation.
①
Press and hold TIMER for about four seconds and then press it
repeatedly until [bSEt] and [mnL] are displayed alternately.
②
Select [StoP] with
△▽
to automatically stop the bath when rotary
evaporator unit completes an operation.
③
Select [KEEP] with
△▽
to automatically retain heat in the bath
when rotary evaporator unit completes an operation.
The factory default setting is "mnL" which disables the bath setting
function.
Set heat retention
temperature
Set heat-retention temperature when "KEEP" is selected. Use
MAINTENANCE to display [Sv2] and enter the desired value for
heat-retention temperature.
①
②
①
②
Содержание BM500
Страница 2: ......