Operation
Multi 3410
51
ba75863y18
01/2015
Note
For
single-point calibration
, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the IDS pH
sensor.
7
Wait for the measurement with stability control to be completed
or terminate the stability control and take over the calibration
value with
<MENU/ENTER>
.
The pH value of the buffer solution is displayed.
8
Set the nominal buffer value for the measured temperature with
<
><
>
.
9
Accept the calibration value with
<MENU/ENTER>
.
The calibration display for the next buffer appears (voltage dis-
play).
10
If necessary, finish the calibration procedure as a single-point
calibration with
<M>
.
The calibration record is displayed.
15.03.2014 08:00
15.03.2014 08:00
Содержание wtw Multi 3410
Страница 1: ...OPERATING MANUAL ba75863e18 01 2015 Multi 3410 DIGITAL METER FOR DIGITAL IDS SENSORS...
Страница 6: ...Multi 3410 6 ba75863y18 01 2015...
Страница 14: ...Multi 3410 Overview 14 ba75863y18 01 2015...
Страница 18: ...Multi 3410 Safety 18 ba75863y18 01 2015...
Страница 22: ...Multi 3410 Commissioning 22 ba75863y18 01 2015 cable 3 3 4 Setting the date and time 3 See section 4 3 4...
Страница 106: ...Multi 3410 Operation 106 ba75863y18 01 2015...
Страница 116: ...Multi 3410 What to do if 116 ba75863y18 01 2015...
Страница 125: ......