Operation
Multi 3410
39
ba75863y18
01/2015
Criteria for a stable
measured value
The
Stability control
function checks whether the measured values are
stable within the monitored time interval.
The minimum duration until a measured value is assessed as stable is
the monitored time interval. The actual duration is mostly longer.
4.6.3
Settings for pH measurements
Overview
The following settings are possible for pH measurements:
Calibration record (display)
Calibration data storage (display)
Calibration interval
Buffers for calibration
Single-point calibration
Unit for slope
Resolution
Settings
The settings are made in the menu for calibration and measurement
settings of the pH/ORP measurement. To open the settings, display the
required parameter in the measured value display and press the
<MENU/ENTER>
key. After completing the settings, switch to the mea-
sured value display with
<M>
.
Default settings are printed in
bold
.
Measured
parameter
Time interval
Stability in the time interval
pH value
15 seconds
Δ
: Better than 0.01 pH
Temperature
15 seconds
Δ
: Better than 0.5 °C
Menu item
Possible setting
Description
Calibration
/
Calibration record
-
Displays the calibration record of the last cal-
ibration
Calibration
/
Calibration data
storage
-
Displays the last calibration records (max.
10)
Calibration
/
Calibration data
storage
/
Output to USB flash
drive or printer
-
Outputs the calibration data storage to the
interface USB-B (USB flash drive/USB
printer)
Calibration
/
Calibration data
storage
/
Output to RS232/USB
-
Outputs the calibration data storage to the
interface USB-B (PC)
Содержание wtw Multi 3410
Страница 1: ...OPERATING MANUAL ba75863e18 01 2015 Multi 3410 DIGITAL METER FOR DIGITAL IDS SENSORS...
Страница 6: ...Multi 3410 6 ba75863y18 01 2015...
Страница 14: ...Multi 3410 Overview 14 ba75863y18 01 2015...
Страница 18: ...Multi 3410 Safety 18 ba75863y18 01 2015...
Страница 22: ...Multi 3410 Commissioning 22 ba75863y18 01 2015 cable 3 3 4 Setting the date and time 3 See section 4 3 4...
Страница 106: ...Multi 3410 Operation 106 ba75863y18 01 2015...
Страница 116: ...Multi 3410 What to do if 116 ba75863y18 01 2015...
Страница 125: ......