Operation
Multi 3410
43
ba75863y18
01/2015
*
Brand names or trade names are trademarks of their respective
owners protected by law.
Note
The buffers are selected in the menu, pH /
<MENU/ENTER>
/
Calibra-
tion
/
Buffer
(see section 4.6.3).
Calibration points
Calibration can be performed using one to five buffer solutions in any
order (single-point to five-point calibration). The meter determines the
following values and calculates the calibration line as follows:
19
Precisa *
3.996
7.003
8.999
25 °C
20
Reagecon TEC
*
2.000
4.010
7.000
10.000
25 °C
21
Reagecon 20
*
2.000
4.000
7.000
10.000
13.000
20 °C
22
Reagecon 25
*
2.000
4.000
7.000
10.000
13.000
25 °C
23
Chemsolute
*
2.000
4.000
7.000
10.000
20 °C
24
USABlueBook
*
4.000
7.000
10.000
25 °C
25
YSI
*
4.000
7.000
10.000
25 °C
No.
Buffer set*
pH values
at