What to do if...
Multi 3410
109
ba75863y18
01/2015
6
What to do if...
6.1
General information
Sensor symbol flashes
Display
Meter does not react to
keystroke
You want to know which
software version is in
the meter or IDS sensor
Cause
Remedy
– Calibration interval expired
– Recalibrate the measuring
system
Cause
Remedy
– Batteries almost empty
– Charge the batteries (see
)
– Replace the batteries (see
)
Cause
Remedy
– Operating condition undefined
or EMC load unallowed
– Processor reset:
Press the
<ENTER>
and
<On/Off>
key simultane-
ously
Cause
Remedy
– E. g., a question by the service
department
– Switch on the meter.
– Open the menu,
<ENTER
_
>
/
Storage &
config
/
System
/
Service
information
. The instrument
data are displayed.
or
– Connect the sensor.
Press softkey [
Info
]/[
More
].
The sensor data are dis-
played (see section 4.5)