Operation
Multi 3410
53
ba75863y18
01/2015
Continuing with three-
to five-point calibration
Note
After the fifth buffer has been measured the calibration is automatically
finished. Then the calibration record is displayed.
The calibration line is determined by linear regression.
19
Thoroughly rinse the IDS pH sensor with deionized water.
20
Immerse the IDS pH sensor in the next buffer solution.
21
When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with
<
><
>
.
22
Start the measurement with
<MENU/ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.
23
Wait for the measurement with stability control to be completed
or terminate the stability control and take over the calibration
value with
<MENU/ENTER>
.
The pH value of the buffer solution is displayed.
24
Set the nominal buffer value for the measured temperature with
<
><
>
.
25
Accept the calibration value with
<MENU/ENTER>
.
The calibration display for the next buffer appears (voltage dis-
play).
26
If necessary, use
<M>
to finish the calibration.
The calibration record is displayed.
or
Continue calibrating using the next buffer with
<MENU/
ENTER>
.
15.03.2014 08:00
Содержание wtw Multi 3410
Страница 1: ...OPERATING MANUAL ba75863e18 01 2015 Multi 3410 DIGITAL METER FOR DIGITAL IDS SENSORS...
Страница 6: ...Multi 3410 6 ba75863y18 01 2015...
Страница 14: ...Multi 3410 Overview 14 ba75863y18 01 2015...
Страница 18: ...Multi 3410 Safety 18 ba75863y18 01 2015...
Страница 22: ...Multi 3410 Commissioning 22 ba75863y18 01 2015 cable 3 3 4 Setting the date and time 3 See section 4 3 4...
Страница 106: ...Multi 3410 Operation 106 ba75863y18 01 2015...
Страница 116: ...Multi 3410 What to do if 116 ba75863y18 01 2015...
Страница 125: ......