Multi 3410
Operation
48
ba75863y18
01/2015
(-59.2 mV/pH at 25 °C) and determines the zero point of the IDS pH
sensor.
Continuing with two-
point calibration
Continuing with three-
to five-point calibration
9
Thoroughly rinse the IDS pH sensor with deionized water.
10
Immerse the IDS pH sensor in buffer solution 2.
11
When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with
<
><
>
.
12
Start the measurement with
<MENU/ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.
13
Wait for the measurement with stability control to be completed
or terminate the stability control and take over the calibration
value with
<MENU/ENTER>
.
The calibration display for the next buffer appears (voltage dis-
play).
14
If necessary, finish the calibration procedure as a two-point cal-
ibration with
<M>
.
The calibration record is displayed.
15.03.2014 08:00
15
Thoroughly rinse the IDS pH sensor with deionized water.
16
Immerse the IDS pH sensor in the next buffer solution.
17
When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with
<
><
>
.
18
Start the measurement with
<MENU/ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.
Содержание wtw Multi 3410
Страница 1: ...OPERATING MANUAL ba75863e18 01 2015 Multi 3410 DIGITAL METER FOR DIGITAL IDS SENSORS...
Страница 6: ...Multi 3410 6 ba75863y18 01 2015...
Страница 14: ...Multi 3410 Overview 14 ba75863y18 01 2015...
Страница 18: ...Multi 3410 Safety 18 ba75863y18 01 2015...
Страница 22: ...Multi 3410 Commissioning 22 ba75863y18 01 2015 cable 3 3 4 Setting the date and time 3 See section 4 3 4...
Страница 106: ...Multi 3410 Operation 106 ba75863y18 01 2015...
Страница 116: ...Multi 3410 What to do if 116 ba75863y18 01 2015...
Страница 125: ......