Multi 3410
Operation
104
ba75863y18
01/2015
The sensor settings are reset under the
Reset
menu item in the menu
for calibration and measurement settings. To open the settings, display
the required parameter in the measured value display and press the
<MENU/ENTER>
key.
Measured parameter
ϰ
Cell constant (c)
Depending on the connected
measuring cell:
0.475 cm
-1
(calibrated)
0.475 cm
-1
(set up)
0.100 cm
-1
Temperature compensation
nLF
Reference temperature
25 °C
Temperature coefficient (TC) of the
linear temperature compensation
2.000 %/K
TDS factor
1,00
Setting
Default settings