SonTek – a Xylem brand
FlowTracker2 User’s Manual (January 2019)
62
SNR Raw Data
4.4.2
The
SNR
variable displays the individual Beam SNR measurements from the one
second measured raw data.
a) SNR is primarily a function of the amount of particulate matter in the water. For
good conditions, SNR should be at least 10 dB. The system can operate
effectively with SNR as low as 2-3 dB, although the noise in individual velocity
measurements will increase,
b) The SNR data is updated once per second,
c) The display options for individual or all beam SNR values are listed in
The shortcut key on the keypad determines which beam SNR will be displayed,
Table 4-4 - Beam SNR Display
Beam
SNR
Line Color
Keypad Key
Beam1
SNR1
Red
1
Beam2
SNR2
Blue
2
Beam3
SNR3
Green
3
All Beams
All SNR
All
4
d) The continuous lines displayed in the graphics are defined under the following
categories,
i).
Thin
continuous line represents the one second raw SNR data,
ii).
Thick
continuous line represents moving average of the one second raw SNR
data.
e) The SNR values displayed in the rectangular boxes for each beam below the
graphical display are defined under the following categories,
i).
Thin
box outline, instantaneous one second raw SNR data,
ii).
Thick
box outline, represents moving average of the one second raw SNR
data.
Temperature Raw Data
4.4.3
The
Temperature
variable displays the temperature measurements from the one second
measured raw data.
a) Temperature data are used for sound speed calculations and can affect velocity
data.
b) Check temperature data to be sure the values are reasonable for the
environment.
c) The Temperature data are updated once per second.
Содержание SonTek FlowTracker2
Страница 107: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 106 ...
Страница 117: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 116 ...
Страница 196: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 195 ii Stationary Measurement ...
Страница 198: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 197 Measurement File Framework 12 1 1 ...
Страница 216: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 215 Variable Mode Description measurement ...
Страница 247: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 246 Configuration Changes ...
Страница 280: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 279 Figure 14 36 Edit by Station Edit a Station ...
Страница 283: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 282 ...
Страница 299: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 298 Appendix A Software Flow Diagram ...
Страница 300: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 299 ...