Std Error Threshold
3.4.3.2
The
Standard Error Threshold
parameter is a maximum threshold placed on the
Standard Error of velocity measurements.
a) Standard error of velocity (
σV
) is a direct measure of the accuracy of the mean
velocity data.
b) The default Standard Error Threshold applied in FlowTracker2 is 0.01m/s.
c)
To change the Standard Error Threshold,
i).
Use the backspace key to clear current entry,
ii).
Type in the new value using the keypad.
d)
To
DISABLE
the Standard Error Threshold,
i).
Use the backspace key to clear current entry until “Disabled” is displayed.
Spike Threshold
3.4.3.3
The
Spike Threshold
parameter is the maximum threshold placed on the percentage of
spikes within a measurement.
a) Spikes in velocity data occur with any acoustic Doppler velocity sensor such as
the FlowTracker. Spikes may have a variety of causes – large particles, air
bubbles, or acoustic anomalies.
b) The default Spike Threshold applied in FlowTracker2 is 10%.
c)
To change the Spikes Threshold,
i).
Use the backspace key to clear current entry,
ii).
Type in the new value using the keypad.
d)
To
DISABLE
the Spikes Threshold,
i).
Use the backspace key to clear current entry until “Disabled” is displayed.
Velocity Angle for Warning
3.4.3.4
The
Velocity Angle for Warning
parameter is the maximum threshold placed on the
velocity angle towards the probe.
a) An angle of 0
°
means flow direction is perpendicular to the tag line (as desired for
an ideal measurement location).
b) The default Velocity Angle for Warning applied in FlowTracker2 is 20 degrees.
c)
To change the Velocity Angle for Warning,
i).
Use the backspace key to clear current entry,
ii).
Type in the new value using the keypad.
d)
To
DISABLE
the Velocity Angle for Warning,
i).
Use the backspace key to clear current entry until “Disabled” is displayed.
Содержание SonTek FlowTracker2
Страница 107: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 106 ...
Страница 117: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 116 ...
Страница 196: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 195 ii Stationary Measurement ...
Страница 198: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 197 Measurement File Framework 12 1 1 ...
Страница 216: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 215 Variable Mode Description measurement ...
Страница 247: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 246 Configuration Changes ...
Страница 280: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 279 Figure 14 36 Edit by Station Edit a Station ...
Страница 283: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 282 ...
Страница 299: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 298 Appendix A Software Flow Diagram ...
Страница 300: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 299 ...