SonTek – a Xylem brand
FlowTracker2 User’s Manual (January 2019)
165
Variable
Description
Units
Decimal
Vel. Z
Mean Z velocity component of all samples.
Velocity is despiked and mounting correction
applied (if applicable).
m/s or ft/s
4
σV. X
Standard error of X velocity component.
m/s or ft/s
4
σV. Y
Standard error of Y velocity component.
m/s or ft/s
4
σV. Z
Standard error of Z velocity component.
m/s or ft/s
4
SNR. 1
Signal to noise ratio of Beam 1
dB
3
SNR. 2
Signal to noise ratio of Beam 2
dB
3
SNR. 3
Signal to noise ratio of Beam 3
dB
3
Angle
Flow angle relative to X direction
Degree
2
Bnd
Boundary QC value
n/a
n/a
Tilt
Mean wading rod angle during point velocity
measurement
Degree
2
Temp
Mean temperature during measurement
°C or °F
2
Review Station Measurement
9.3.2.3
Review Station
measurement is a tabular report that is available from the data collection
window at the completion of point velocity measurement of a station.
Review Station
screen consists of tabular summary
of measurements performed at the station,
b)
To access review station report,
i).
Select the station in the data collection window,
ii).
Press the enter key.
c)
To
EXIT
review station report,
i).
Press the Left Soft Key.
Figure 9:16 - Review Station
Properties
associated with the variables and numerical calculations performed on the
raw data collected at the station is defined in
Table 9-5 - Review Station
Variable
Description
Units
Decimal
Loc X
Location X coordinate
m or ft
3
Loc Y
Location Y coordinate
m or ft
3
Depth
Measured water depth
m or ft
3
Meas. Depth
Velocity measurement depth
m or ft
3
Comment
Comment supplied before the measurement
n/a
n/a
Warning
Warning supplied based on quality control
criteria
n/a
n/a
Samples
Total number of sample received
number
n/a
Spikes
Number of spikes removed from mean velocity
number
n/a
Содержание SonTek FlowTracker2
Страница 107: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 106 ...
Страница 117: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 116 ...
Страница 196: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 195 ii Stationary Measurement ...
Страница 198: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 197 Measurement File Framework 12 1 1 ...
Страница 216: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 215 Variable Mode Description measurement ...
Страница 247: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 246 Configuration Changes ...
Страница 280: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 279 Figure 14 36 Edit by Station Edit a Station ...
Страница 283: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 282 ...
Страница 299: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 298 Appendix A Software Flow Diagram ...
Страница 300: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 299 ...