![Xylem SonTek FlowTracker2 Скачать руководство пользователя страница 176](http://html1.mh-extra.com/html/xylem/sontek-flowtracker2/sontek-flowtracker2_user-manual_3888944176.webp)
SonTek – a Xylem brand
FlowTracker2 User’s Manual (January 2019)
175
Step 1
PS Calibration
screen allows the user to perform
pressure sensor calibration. The user is required to
remove the probe from the water.
a)
To perform PS Calibration,
i).
Press the Right Soft Key.
b)
To
CANCEL
PS Calibration,
i).
Press the Left Soft Key.
Figure 8:29 – PS calibration
Step2
PS Calibration
screen with
Quality Control
warning
messages based on quality control checks performed
on measurement data.
a)
To accept quality control warning messages,
i).
Press the Right Soft Key.
b)
To
REDO
PS Calibration,
i).
Press the Left Soft Key.
Figure 8:30 - Warning Messages
Complete Measurement
9.3.4.5
The
Complete Measurement
function closes the measurement file and the user will not
be able to add additional stations and or make any changes to the data captured. The
complete measurement function is a three-step process, firstly to determine if the user is
satisfied with the velocity measurements, secondly to perform quality control checks
against all measurement data and lastly to review measurement summary.
The Complete Measurement process allows the user to review the measurement data
before the complete function is selected at Summary screen. The user will not be able
to make any to changes to the measurement file from the handheld when this function
is selected.
Step 1
Confirmation
screen of
Complete Measurement
function confirms if the user wants to close the
measurement file. User will not be able to add
stations or make any changes to the data
a)
To confirm complete measurement operation,
i).
Press the Right Soft Key.
b)
To
CANCEL
complete measurement operation,
i).
Press the Left Soft Key.
Figure 9:31 - Confirmation
Содержание SonTek FlowTracker2
Страница 107: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 106 ...
Страница 117: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 116 ...
Страница 196: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 195 ii Stationary Measurement ...
Страница 198: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 197 Measurement File Framework 12 1 1 ...
Страница 216: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 215 Variable Mode Description measurement ...
Страница 247: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 246 Configuration Changes ...
Страница 280: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 279 Figure 14 36 Edit by Station Edit a Station ...
Страница 283: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 282 ...
Страница 299: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 298 Appendix A Software Flow Diagram ...
Страница 300: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 299 ...