![Xylem SonTek FlowTracker2 Скачать руководство пользователя страница 46](http://html1.mh-extra.com/html/xylem/sontek-flowtracker2/sontek-flowtracker2_user-manual_3888944046.webp)
SonTek – a Xylem brand
FlowTracker2 User’s Manual (January 2019)
45
d) Salinity should be specified or determined
using a CTD instrument as accurately as
possible for each measurement location. The
CastAway-CTD
is a lightweight, easy to use
instrument designed for quick and accurate
conductivity, temperature, and depth profiles.
Starting with a unique six-electrode
conductivity cell and fast response thermistor
the CastAway makes use of modern
technology to provide state of the art CTD
measurements.
Figure 3:12 - CastAway - CTD
e) Sound speed is used in Doppler velocity calculations. See the
for details about the effect of sound speed on velocity data.
f)
To change the Salinity,
i).
Use the backspace key to clear current entry,
ii).
Type in the new value using the keypad.
Temperature
3.4.2.3
The
Temperature
parameter specifies the temperature value used to compute sound
speed.
a) The temperature value used to compute sound speed is based on internal
temperature sensor by default. The temperature field will be populated with the
value “
Measured
” to indicate it is based on internal temperature sensor.
b) Water temperature measured by external devices can be entered in the
temperature field. The temperature value from the external sensor will then be
used for sound speed calculations.
The temperature field cannot be used to store reference temperature readings as this
will be used in the sound speed calculations. It is recommended that reference
temperature readings be documented in the comments fields.
c) Sound speed is used in Doppler velocity calculations. See the
for details about the effect of sound speed on velocity data.
d)
Enter external temperature reading,
i).
If the default value “Measured” is displayed, start typing using keypad, otherwise,
ii).
Use the backspace key to clear current entry,
iii).
Type in the new value using the keypad.
Sound Speed
3.4.2.4
The
Sound Speed
parameter specifies the sound speed value used in Doppler velocity
calculations.
Содержание SonTek FlowTracker2
Страница 107: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 106 ...
Страница 117: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 116 ...
Страница 196: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 195 ii Stationary Measurement ...
Страница 198: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 197 Measurement File Framework 12 1 1 ...
Страница 216: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 215 Variable Mode Description measurement ...
Страница 247: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 246 Configuration Changes ...
Страница 280: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 279 Figure 14 36 Edit by Station Edit a Station ...
Страница 283: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 282 ...
Страница 299: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 298 Appendix A Software Flow Diagram ...
Страница 300: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 299 ...