Handheld Maintenance

Wipe the keypad, screen, and case with a cloth dampened with a mild solution 
of clean water and dish soap. Optimal storage temperature of the handheld 
instrument is 0-45°C. The battery pack permanently loses capacity at a faster rate 
when above 45°C.

Short-term Storage:

 

Assure that the handheld instrument is powered off, and store it in a temperature-
controlled, secure location. Ideally all ports should be covered to prevent dust, 
water, or other contamination.

Long-term Storage:

 

In addition to the short-term storage guidelines above, remove the battery pack 
to prevent any damage from possible battery leaks. Reinstall the battery cover. 
Store the battery pack in a dry place ideally around 25°C.

  

Sensors, Calibration, and Maintenance

Содержание ProSwap 610224-01

Страница 1: ...ProSwap Logger User Manual 1 PORT TEMPERATURE DEPTH SONDE USER MANUAL ITEM 610224 01 REVISION A...

Страница 2: ...l info ysi com YSI com Safety Information Please read this entire manual before unpacking setting up or operating this equipment Pay attention to all precautionary statements Failure to do so could re...

Страница 3: ...on Setup 4 2 Calibration Screen 4 3 Temperature Sensor 4 4 Depth Sensor 4 5 Conductivity Sensor 4 6 Dissolved Oxygen Sensor 4 7 pH and pH ORP Sensors 4 8 Ion Selective Electrodes 4 9 Turbidity Sensor...

Страница 4: ...4 Section 1 Introduction This section will cover the ProSwap Logger specifications sonde contents power supply and introduce some relevant accessories...

Страница 5: ...ogger via the ProSwap or ProDSS Handhelds or via Kor Software with a USB communications adapter to a PC Specifications Material Titanium Internal Logging Memory Capacity 512 MB 100 000 data sets Sampl...

Страница 6: ...6 Sensor Probe Guard Cable Connector Vent Tube Cable Depth Sensor Thermistor Battery Compartment Sensor Port ProSwap Logger...

Страница 7: ...ts ProSwap Logger with Integrated Cable Probe Guard Weight for Probe Guard Storage Sleeve Sponge Graduated Cylinder USB Drive contains Kor Software Cable Connector Cap Cable Grip Kit Cable Management...

Страница 8: ...connector into the AC power adapter or a computer USB connector to maintain charging Alternatively the battery can be charged using the USB adapter accessory sold separately Plug the USB cable into t...

Страница 9: ...nd operation of the ProSwap Logger using a ProDIGITAL Handheld ProSwap and ProDSS To prepare the sonde for operation you will need to Install sensor on the bulkhead or plug the port Connect ProSwap Lo...

Страница 10: ...low any ProDSS smart sensor to be connected NOTICE The sensor connector and bulkhead port are not wet mateable Make sure that the sensor connector and bulkhead port are clean and dry before sensor ins...

Страница 11: ...sor installation removal tool into the retaining nut then rotate the retaining nut counterclockwise to loosen 2 After the retaining nut has been completely unscrewed from the bulkhead pull the sensor...

Страница 12: ...ly to prevent cross threading Incorrect installation may cause damage to the sensor guard or bulkhead that is not covered by the warranty To help stabilize the sensors when profiling at deeper depths...

Страница 13: ...the keys on the cable connector with the slots on the handheld connector 2 Push together firmly and then twist the outer ring clockwise until it locks into place 3 Power on the handheld to calibrate...

Страница 14: ...install 4 After installing Kor go back to the main installer window and select Install USB Drivers 5 Select Install All and follow the prompts to install 6 Connect the PSL to a USB Adapter or handheld...

Страница 15: ...attended deployment perform sensor calibrations and more Users may consult the ProDIGITAL User Manual for additional information about the handheld meter and ProDSS Sensors The follow text will be rel...

Страница 16: ...rate Opens the calibration menu in order to calibrate sensors or restore the factory default calibration 9 Help Shows context sensitive help 4 File Opens the file menu in order to view delete and back...

Страница 17: ...to the Run screen To enable or disable an option highlight the option then push the ENTER key Enabled functions appear as a circle with a dot or a box with a check mark Disabled functions appear as a...

Страница 18: ...essage area shows status messages error messages and information about selected functions 1 Date Time 7 Sonde Deployed Indicator 2 GPS Signal Indicator 8 USB PC Connection Indicator 3 Handheld Battery...

Страница 19: ...to Stable 4 Verify that the sensor and sensor guard are correctly installed 5 Insert the probe into the sample NOTE NOTE Make sure to submerge the sensors completely 6 Move the bulkhead of the sonde i...

Страница 20: ...ions Set YY MM DD MM DD YY DD MM YY or YY DD MM date format Set the correct date Select 12 or 24 hour time format Set the correct time 1 5 9 3 7 11 13 2 6 10 4 8 12 1 Set the Date and Time 2 Change th...

Страница 21: ...l memory can save up to 400 individual calibration records After 400 records the instrument will overwrite previously stored calibration records starting with the oldest To prevent the permanent loss...

Страница 22: ...he password in a safe place Contact YSI Technical Support if you lose the password Technical support Language Language The instrument is shipped with English enabled If a different language is desired...

Страница 23: ...anumeric order or Distance from the current position Choose an entry from the Site List or Data ID List to Select Edit or Delete When selected data recorded will be tagged with the specific site and o...

Страница 24: ...measurements for 15 seconds before logging a data set Auto Shutoff Auto Shutoff To conserve battery power auto shutoff powers off the instrument after a user defined time period in minutes The auto sh...

Страница 25: ...nt Note the serial number when contacting YSI support Unit ID Unit ID Users can set a custom Unit ID The Unit ID identifies the instrument in Kor Software Sensor Info Sensor Info Sensor info shows mea...

Страница 26: ...can measure virtual vented depth The virtual vented depth measurement allows for real time compensation for atmospheric pressure using the handheld s barometer Depth offset Depth offset can be used i...

Страница 27: ...the nature of the ionic species present in the water sample To be assured of moderate accuracy for the conversion you must determine a multiplier for the water at your sampling site Use the following...

Страница 28: ...ecognition 4 00 7 00 and 10 00 or NIST auto buffer recognition 4 01 6 86 and 9 18 Calibration values are automatically compensated for temperature for both buffer sets Setup Turbidity Setup Turbidity...

Страница 29: ...when the measurement is stable Select a sensor to enable or disable Auto Stable Then set the stability threshold parameters The Auto Stable stability threshold can be set by percent of measurement or...

Страница 30: ...surement is detected allowing the handheld to adapt when an event occurs The Default mode provides optimum averaging for all sensors This mode has up to 40 seconds of averaging on some sensors to curb...

Страница 31: ...h the warranty period of the ODO Cap ProDSS ODO Sensor Cap SKU 626890 12 months The handheld will automatically recognize the last time the ODO Sensor Cap coefficients were updated and alert the user...

Страница 32: ...ata Select Now to immediately begin autonomously data collection or select Next Interval to begin collecting data on the next logging interval Select Custom Time to have the sonde begin data collectio...

Страница 33: ...tput menu is used to configure SDI 12 output settings for users that are connecting their ProSwap Logger to an external data logger NOTE A Flying Lead Adapter is needed to connect the ProSwap Logger t...

Страница 34: ...th without waiting for the depth data to stabilize Rapid mode should be used when the sonde is moving quickly through the water such as with rapid profiling and unique applications e g towed applicati...

Страница 35: ...35 Deploy Status Select Status to view the sonde information and check if logging is enabled or disabled...

Страница 36: ...the Calibration section for sensor specific calibration procedures NOTE User ID Probe ID and User Field 1 and 2 can be enabled in the Calibration Settings under the System menu 1 Sensors connected 4 R...

Страница 37: ...y Use the Transfer Sonde Data menu to transfer sonde data to the Handheld Enter the desired filter criteria then select Transfer Sonde Data to transfer data from the sonde NOTE Make sure the date rang...

Страница 38: ...me ranges View Cal Records View Cal Records Use this menu to view all calibration records stored in memory Use the and arrow keys to scroll through different calibration records Calibration informatio...

Страница 39: ...ia then connect the handheld to the USB storage device using the supplied adapter Select Backup Data to export the data to an USB storage device NOTE The USB storage device must be formatted as FAT32...

Страница 40: ...Logger and ProDIGITAL Sensors except temperature require periodic calibration Calibration procedures follow the same basic steps with variations for specific parameters Before performing a calibration...

Страница 41: ...n extended guard Item 626740 is sold separately and required for use with Turbidity and Total Algae sensors For highest data accuracy thoroughly rinse the graduated cylinder guard and sensors with a s...

Страница 42: ...e Observe the White Line to ensure the measurement is stable before choosing Accept Calibration Post Cal Value This is the same as the calibration value This will be the measurement value in the curre...

Страница 43: ...epth at the location of measurement A change in barometric pressure will result in a zero shift for non vented sensors unless the transducer is recalibrated to the new pressure 1 Make sure that the de...

Страница 44: ...rature Operating Storage 5 to 50 C 20 to 80 C Accuracy 5 to 50 C 0 25 C Response T63 1 sec Resolution 0 1 C Sensor Type Thermistor The conductivity temperature sensor can measure and calculate conduct...

Страница 45: ...mS Make sure that the units are correct and match the units displayed on the handheld 7 Observe the actual measurement readings for stability white line on graph shows no significant change for 40 se...

Страница 46: ...re is no oxygen present the lifetime of the signal is maximal as oxygen is introduced to the membrane surface of the sensor the lifetime becomes shorter Thus the lifetime of the luminescence is invers...

Страница 47: ...Temperature Operating Storage 5 to 50 C 20 to 80 C Range 0 to 500 air sat 0 to 50 mg L Accuracy 0 200 1 reading or 1 air sat whichever is greater 200 500 5 reading 0 20 mg L 1 of reading or 0 1 mg L 2...

Страница 48: ...ce the sensor cap as needed Dissolved Oxygen Calibration Optical Dissolved Oxygen ODO calibration requires the current true barometric pressure Make sure that the barometer is read ing accurately prio...

Страница 49: ...he bulkhead and sensors in clean water then dry 5 Perform a ODO water saturated air calibration after performing a zero point calibration Dissolved Oxygen Sensor Maintenance The ODO sensor should be k...

Страница 50: ...sponge if it becomes dirty A grey storage sleeve is shipped with the ProSwap Logger for an easy storage option Simply moisten the sponge with a small amount of clean water and slide the sleeve over t...

Страница 51: ...event scratches Do not use organic solvents to clean the ODO sensor lens 6 Remove the new sensor cap from its hydrated container and dry the inside cavity of the sensor cap with lens cleaning tissue M...

Страница 52: ...Push the ENTER key after each entry and then proceed to the next K selection 4 After all the new coefficients have been entered select Update Sensor Cap Coefficients 5 A message will appear warning t...

Страница 53: ...posed to the sample where the concentration of hydrogen ions varies The resulting differential creates a potential read by the meter versus the stable potential of the reference Specifications pH Unit...

Страница 54: ...H mV readings during calibration to understand the condition and response of the pH sensor In buffer 7 pH mVs should be between 50 and 50 In pH4 buffer the mV reading should be 165 to 180 mV higher th...

Страница 55: ...lution it doesn t matter which one comes next 8 Immerse the probe into the buffer solution The Calibration value will automatically be adjusted based on the selected buffer and temperature 9 Wait for...

Страница 56: ...ration and press the ENTER key Calibration successful will be displayed in the message area Sensors Calibration and Maintenance pH ORP Sensor Maintenance The pH and pH ORP sensors are shipped with the...

Страница 57: ...sed from most chemical or laboratory distributors To prevent injury carefully follow the HCl manufacturer s instructions If HCl is not available soak in white vinegar 2 After soaking thoroughly rinse...

Страница 58: ...torage bottle features an open cap and o ring to form a tight seal around the sensor tip the solution contains KCl with potassium phthalate and a preservative If this original solution is not availabl...

Страница 59: ...rocking slightly if necessary Do not keep the used o rings as they are unusable after removal from the sensor body Discard the old sensor module 5 Inspect the sensor connector port for debris or mois...

Страница 60: ...ns When the sensor is immersed in water a potential is established across the membrane that depends on the relative amounts of ions in the sample and the internal solution This potential is read relat...

Страница 61: ...ion solution Due to the solid state nature of the chloride ISE the sensor exhibits moderate flow dependence Mitigation can be achieved by stirring during calibration ISE Precautions ISEs should only b...

Страница 62: ...magnitude 1 mg L and 100 mg L for Ammonium and Nitrate 10 mg L and 1 000 mg L for Chloride 1 Fill the graduated cylinder to the appropriate level standard for calibration point 1 Immerse the probe in...

Страница 63: ...00 mL glass or plastic storage vessels CAUTION Hydrochloric acid is highly corrosive and toxic and should therefore be handled with extreme care in a well ventilated fume hood The equivalent amount of...

Страница 64: ...f anhydrous potassium nitrate and transfer quantitatively into a 1000 mL volumetric flask Add 1 0 g of anhydrous magnesium sulfate to the flask 2 Add approximately 500 mL of water to the flask Swirl t...

Страница 65: ...swirl to dissolve the solid reagents then dilute to the volumetric mark with water 4 Mix well by repeated inversion then transfer the 10 mg L standard to a storage bottle Sensors Calibration and Main...

Страница 66: ...ours preferably overnight in the sensor s high calibration solution Sensor Module Ammonium chloride and nitrate sensors feature user replaceable sensor modules These modules contain a reference soluti...

Страница 67: ...on 860 15 nm Turbidity Sensor 4 9 Turbidity is the indirect measurement of the suspended solid concentration in water and is typically determined by shining a light beam into the sample solution and t...

Страница 68: ...d according to details in Standard Methods for the Treatment of Water and Wastewater Section 2130 B These standards include YSI Certified AMCO AEPA polymer based standards see above Hach StablCal stan...

Страница 69: ...displayed in the message area 5 Discard the used standard and rinse the probe guard and graduated cylinder with a small amount of the next calibration point standard Discard the rinse standard 6 Fill...

Страница 70: ...U g L PC RFU g L PE Temperature Operating Storage 5 to 50 C 20 to 80 C Range Chl 0 100 RFU 0 400 g L Chl PC 0 100 RFU 0 100 g L PE 0 100 RFU 0 280 g L Response T63 2 sec Resolution Chl 0 01 RFU 0 01 g...

Страница 71: ...and an environmental algal population can behave quite differently This is also why the TAL sensors and in situ monitoring should not be regarded as a perfect replacement for other methods such as pig...

Страница 72: ...solution of Rhodamine WT Transfer 5 0 mL of the 125 mg L solution prepared in step one into a 1000 mL volumetric flask Fill the flask to the volumetric mark with deionized or distilled water Mix well...

Страница 73: ...ing of the sensor This calibration does not reset the second point entered during the previous 2 point calibration The consequence is that error will be alleviated at and near zero but more error can...

Страница 74: ...graph shows no significant change for 40 seconds and then select Accept Calibration Ready for cal point 2 will be displayed in the message area 6 Discard the used water and rinse the probe guard and g...

Страница 75: ...ree cloth If necessary use mild soapy water The sensor can be stored wet or dry For long term storage YSI recommends storing the sensor dry Install the shipping cap or sensor guard to prevent scratche...

Страница 76: ...e uncorrected Use this approximate formula True BP in mmHg Corrected BP in mmHg 2 5 Local altitude in ft above sea level 100 Example Corrected BP 759 mmHg Local altitude above sea level 978 ft True BP...

Страница 77: ...rt term Storage Assure that the handheld instrument is powered off and store it in a temperature controlled secure location Ideally all ports should be covered to prevent dust water or other contamina...

Страница 78: ...ers to view record live data view transfer recorded data calibrate sensors set up the logger for deployment and more Kor Software and drivers require permissions for successful installation Administra...

Страница 79: ...Basic Microsoft Windows 8 Home Premium Microsoft Windows 8 Professional Microsoft Windows 8 Enterprise Microsoft Windows 8 1 Basic Microsoft Windows 8 1 Professional Microsoft Windows 8 1 Enterprise...

Страница 80: ...flash drive in Windows Explorer right click on Start exe and select Run as administrator to start the installer 3 On the Kor Installer click Install Driver Then choose to Install the driver on the sc...

Страница 81: ...Connect the ProSwap Logger to a USB adapter or handheld 3 Connect the USB cable to the PC Click Connect in the Instrument Connection Panel to establish communication 4 Check for firmware updates If a...

Страница 82: ...ted 4 ProDIGITAL Handheld Allows users to configure handheld specific settings 11 Home Screen Provides quick access to the most common functions of the software and links to helpful pages on YSI com 5...

Страница 83: ...ngs related to the software as well as parameter specific settings It is important to note that these settings are saved locally and only pertain to the software itself These settings are not pushed t...

Страница 84: ...spective sensor serial number STARTUP OPTIONS Require User Login Toggle On Off This requires the user to select a User Name when the software is launched The selected User Name will be the default ID...

Страница 85: ...ote that these settings are saved locally to Kor Software and do not change sensor hardware settings NOTE If a parameter is disabled then it will not show up in the calibration menu About Users can vi...

Страница 86: ...to restore the factory default calibration Note this deletes the user calibrations from the sensor and reverts to the original factory settings A user calibration must be performed after the factory r...

Страница 87: ...e Calibration Records Panel Manage Sensor Reminders Reminders may be enabled or disabled for select parameters based on a predefined calibration interval This interval may be adjusted by the user Addi...

Страница 88: ...ProDIGITAL ProDSS ProSwap and ProSolo Handheld meter These settings will only be applied to a connected handheld and are handheld specific Configuration files may be read from a connected handheld and...

Страница 89: ...the next time interval as specified by the deployment template 2 NOW Logging will begin immediately 3 CUSTOM Logging will begin at a user specified date and time Deployment Template A deployment templ...

Страница 90: ...e sonde will log data File Name Prefix this is the file name under which the logged data will be saved Site Name name of the location to be tagged with the logged data User Name name of the user to be...

Страница 91: ...e SDI 12 Parameters all parameters available to select and organize Selected SDI 12 Parameters this dictates which parameters the sonde will send to the SDI 12 data logger and in what order Move param...

Страница 92: ...ault mode In Accelerated mode sensors record data with a smaller rolling average window 5 10 seconds so changes in sensor response are more quickly observed Accelerated mode is recommended when the se...

Страница 93: ...sampling Kor Software Create a New Site Users can input a custom site name required and a site description optional The site creation date is auto populated Additional options include adding a site p...

Страница 94: ...rt Saving Data Logs continuously at the specified time interval Stop Saving Data Stops the continuous logging Current Site The active site that is tagged to the logged data Interval The time interval...

Страница 95: ...hich are refreshed at the specified time interval GRAPH A time based or depth based graph view each graph can display up to two parameters specified by the user TABLE A column based view where new row...

Страница 96: ...wed in Table or Graph view Additionally data can be exported or printed Search Access and filter the software database to find logged data files multiple files can be viewed simultaneously Export to C...

Страница 97: ...mported from the sonde or handheld Users can also delete data files that are saved on connected instruments To delete data click the Delete Records button under the Logged Measurement Files box Select...

Страница 98: ...98 Section 6 Accessories This section will cover the common accessories designed for use with the ProSwap Logger...

Страница 99: ...rated Color LCD graphic display Memory 100 000 data sets Software Kor Software Communications Field Cable USB Power Internal Operating Time Charging Time Rechargeable Lithium Ion Pack 15 hours 9 hours...

Страница 100: ...ment for 6 AA alkaline batteries Twist lock Cable Connector Battery Level Indicator Magnetic Switch to check Battery Level Specifications Batteries 6 AA Alkaline Battery Life At least 90 days at 15 mi...

Страница 101: ...AA AA AA AA AA AA AA AA AA AA AA AA AA AA AA AA Setting Up 1 The Power Pack cover utilizes captive screws Use a Phillips head screwdriver to loosen the 4 screws and remove the cover to access the batt...

Страница 102: ...ication between the ProSwap Logger and Kor Software It also includes a power adapter for AC charging 1 2 3 Software Interface 1 Connect the ProSwap Logger cable to the USB Adapter 2 Connect the USB mi...

Страница 103: ...allows for AC charging of the ProSwap Logger 1 Connect the ProSwap Logger cable to the USB Adapter 2 Connect the USB micro B plug of the cable to the USB Adapter 3 Connect the USB Type A plug of the c...

Страница 104: ...facilitates attachment of the cartridge to the barb fitting of the cable and a cap which protects the MS 8 connector of the cable during deployment Installing the Cartridge Refer to the following dra...

Страница 105: ...e Flying Lead Adapter Accessories Mounting Users may desire to mount the Flying Lead Adapter in an enclosure Please refer to the dimensional diagram for reference Cable Conductor Connector Pin Number...

Страница 106: ...106 Section 7 Signal Output The ProSwap Logger can output to a DCP via the Flying Lead Adapter This section covers the setup and configuration to output via SDI 12 communication protocol...

Страница 107: ...them using the arrow buttons See Section 5 8 DCP Adapter Output for more information Once the parameter output set save and apply the template to the sonde This template can be saved locally on the PC...

Страница 108: ...fication M Start Measurement V Start Verification Extended Commands Supports the following extended commands Zsn Returns the manufacturer s serial number of the ProSwap Logger Zpara Returns the SDI 12...

Страница 109: ...dity NTU 37 Parameter Code NH4 Ammonium mg L 48 Date DDMMYY 51 Date MMDDYY 52 Date YYMMDD 53 Time HHMMSS 54 TDS kg L 95 NO3 Nitrate mV 101 NO3 Nitrate mg L 106 NH4 Ammonium mV 108 TDS mg L 110 Chlorid...

Страница 110: ...tion 8 Safety and Support This section will cover information related to safety and precautions while using the ProSwap Logger and accessories as well as contact information for service and technical...

Страница 111: ...y metal object e g wire DO NOT carry or store the battery pack with necklaces hairpins or other metal objects DO NOT carry or store the battery pack with hazardous or combustible materials DO NOT pier...

Страница 112: ...F to 140 F Use of the battery pack outside of this temperature range may damage the performance of the battery pack or may reduce its life expectancy DO NOT discharge the battery pack using any device...

Страница 113: ...ally For the nearest service center information please visit YSI com highlight Customer Support and then click Product Service or contact YSI Technical Support directly at 800 897 4151 1 937 767 7241...

Страница 114: ...610150 x 610151 x 610152 x 610153 x 610154 x 610155 x Accessories 610175 610185 006108 627195 626740 696162 Conforms to the following Directives EMC 2014 30 EC LVD 2014 35 EU WEEE 2012 19 EU RoHS 2011...

Страница 115: ...or visit www YSI com Support tab for a Product Return Form Send the product and proof of purchase transportation prepaid to the Authorized Service Center selected by YSI Repair or replacement will be...

Страница 116: ...tral to our work Our products and services move treat analyze monitor and return water to the environment in public utility industrial residential and commercial building services settings Xylem also...

Отзывы: