Long-term Storage:
• Method 1:
Submerge the sensing end of the sensor in a container of distilled
or deionized water. Periodically check the level of the water to make sure that it
does not evaporate.
• Method 2:
Wet the sponge located in the cap originally included with the
ODO sensor, then install on sensing end of the ODO sensor. Replace the
sponge if it becomes dirty.
A grey storage sleeve is shipped with the ProSwap Logger for an easy storage
option. Simply moisten the sponge with a small amount of clean water and slide
the sleeve over the probe guard to create a moist atmosphere for the sensor.
ODO Sensor Rehydration
If the ODO sensor has accidentally been left dry for longer than 8 hours, it must
be rehydrated. To rehydrate, soak the ODO sensor in room temperature tap water
for approximately 24 hours. After the soak, calibrate the sensor.
Sensors, Calibration, and Maintenance