21
Operation
Calibration Options
→
Calibration Record
→
Options
User ID, Probe ID, or User Field #1 or 2 can be user-defined for positive calibration
file identification of:
• The person calibrating the instrument
• The sensor/cable serial number used during calibration (or other,
user-defined Probe ID)
• Other user-specific identification (User Field #1 and #2)
NOTE:
User Field can be used to describe the condition of the probe. For
example, new sensor or new ODO cap.
Calibration Record
Detailed sensor calibration information is stored for later review. The instrument’s internal memory can save up to 400 individual
calibration records. After 400 records, the instrument will overwrite previously stored calibration records, starting with the oldest. To
prevent the permanent loss of calibration records, periodically download the calibration files to a computer using the Kor Software.
Re-Cal Prompts
→
Calibration Record
→
Options
→
Re-Cal Prompts
Re-Cal Prompts provide a reminder to recalibrate a probe in the
user-defined number of days. Select the desired sensor Re-Cal prompt, then enter
the desired number of days before the Re-Cal prompt occurs. This reminder will
be provided when the instrument is powered on and will reoccur every day until
the sensor is re-calibrated.
Set the sensor value to zero (0) days (default) to turn off Re-Cal prompts.