33
Operation
→
Deploy
→
Mode
This advanced menu allows users to customize how the ProSwap Logger sonde
logs data.
Sample & Hold:
This mode is useful for when users would like to log data to the
sonde’s memory and also output data via SDI-12. This mode allows the sonde
to log internally at its log interval and when a data logger sends a measure
command to the sonde, it will respond back with the last internally logged value
instead of taking a new measurement.
Normal:
The sonde will collect and average data (based on the averaging mode)
into a single data point at the user-specified interval.
→
Deploy
→
DCP Adapter Output 1&2
The DCP (Data Collection Platform) Adapter Output menu is used to configure
SDI-12 output settings for users that are connecting their ProSwap Logger to an
external data logger.
NOTE:
A Flying Lead Adapter is needed to connect the ProSwap Logger
to a data logger.
Set the address to a number (0-9) or a single letter (A-Z or a-z) so that the data
logger can identify from which source the data should be pulled.
Set which parameters the data logger will record and the order in which that data
is collected.
NOTE:
The sonde and the data logger must be set with the same
parameters, in the same order.
→
Deploy
→
Sonde Settings
View or change sonde-specific settings in the Sonde Settings menu, including the
sonde date and time, and averaging mode.
Use the Sonde Settings menu to:
• View or change the Sonde ID
• Set the sonde date and time
NOTE:
YSI recommends that users
Sync Time with handheld
prior to
starting a deployment.